FDA Adverse Event Malfunction Summary report: N

BINAXNOW¿ COVID-19 ANTIGEN SELF TEST

MDR report key: 15426942 · Received September 15, 2022

Report

Report Number
1221359-2022-04446
Event Type
Malfunction
Date Received
September 15, 2022
Date of Event
September 4, 2022
Report Date
September 16, 2022
Manufacturer
ABBOTT DIAGNOSTICS SCARBOROUGH, INC.
Product Code
QKP
UDI-DI
00811877011408
PMA / PMN Number
EUA210264
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

THE INVESTIGATION IS STILL IN PROGRESS. A SUPPLEMENTAL REPORT WILL BE PROVIDED AFTER COMPLETION.

Additional Manufacturer Narrative · 0

THIS SUPPLEMENTAL REPORT IS BEING SUBMITTED TO PROVIDE THE INVESTIGATION CONCLUSION. TESTING WAS PERFORMED AT ABBOTT DIAGNOSTICS SCARBOROUGH, INC. ON RETAINED KIT LOT 205787 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR TEST KIT PART NUMBER 195-160 / LOT 205787 AND TEST BASE PART NUMBER 195-430H / LOT 201382. THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. A REVIEW OF THE COMPLAINTS REPORTED AS FALSE NEGATIVE PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) RELATED TO KIT LOT 205787 SHOWED THAT THE COMPLAINT RATE IS (B)(4). ABBOTT DIAGNOSTICS SCARBOROUGH WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE; HOWEVER, IT COULD POSSIBLY BE RELATED TO ISSUES INCLUDING THE SELF-TEST USER PERFORMANCE, INTERPRETATION OF THE RESULT, OR THE SPECIFIC PATIENT SAMPLE.

Description of Event or Problem · 0

THE CONSUMER REPORTED FALSE NEGATIVE RESULTS WITH THE BINAX NOW COVID-19 ANTIGEN SELF TEST PERFORMED ON (B)(6) 2022. ON THE SAME DAY THE CONSUMER PERFORMED A TEST (UNKNOWN PLAFORM AND BRAND) AT A PHARMACY AND GENERATED POSITIVE RESULTS. NO ADDITIONAL PATIENT INFORMATION, INCLUDING TREATMENT AND OUTCOME, WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
428025 BINAXNOW¿ COVID-19 ANTIGEN SELF TEST CORONAVIRUS ANTIGEN DETECTION TEST SYSTEM QKP ABBOTT DIAGNOSTICS SCARBOROUGH, INC. 205787 00811877011408

Patients

Seq Age Sex Outcome Treatment
1 Unknown