FDA Adverse Event Injury Summary report: N

GAP SPACER

MDR report key: 15426458 · Received September 15, 2022

Report

Report Number
1020279-2022-04103
Event Type
Injury
Date Received
September 15, 2022
Date of Event
August 23, 2022
Report Date
October 12, 2022
Manufacturer
SMITH & NEPHEW, INC.
Product Code
HWT
UDI-DI
00885556578445
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MD, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

INTERNAL COMPLAINT REFERENCE: (B)(4).

Additional Manufacturer Narrative · 0

THE ASSOCIATED DEVICE WAS RETURNED AND EVALUATED. A VISUAL INSPECTION OF THE RETURNED DEVICE CONFIRMED THE STATED FAILURE MODE. THE DEVICE FRACTURED IN HALF, RENDERING THE DEVICE INOPERATIVE. A REVIEW OF COMPLAINT HISTORY REVEALED SIMILAR EVENTS FOR THE LISTED DEVICE OVER THE PREVIOUS 12 MONTHS, BUT NO SIMILAR EVENTS FOR THE BATCH BASED ON THE HISTORICAL DATA, THIS FAILURE MODE WILL BE MONITORED FOR FUTURE COMPLAINTS FOR ANY NECESSARY CORRECTIVE ACTIONS. A REVIEW OF THE MANUFACTURING RECORDS DID NOT REVEAL A MANUFACTURING ABNORMALITY THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED INCIDENT. AT THIS TIME, WE DO NOT HAVE REASON TO SUSPECT THAT THE PRODUCT FAILED TO MEET ANY PRODUCT SPECIFICATIONS AT THE TIME OF MANUFACTURE. A REVIEW OF THE RISK MANAGEMENT FILE REVEALED THIS FAILURE MODE WAS PREVIOUSLY IDENTIFIED. THE ANTICIPATED RISK LEVEL IS STILL ADEQUATE. ASSESSMENT OF HISTORICAL ESCALATED CASES CONCLUDED THAT THERE ARE NO PRIOR ACTIONS RELATED TO THIS DEVICE AND FAILURE MODE. AS THE DEVICE BROKE AND IT CANNOT LONGER FIT ITS PURPOSE, THE CONTRIBUTION OF THE DEVICE TO THE REPORTED EVENT COULD BE CORROBORATED. THIS DEVICE IS A REUSABLE INSTRUMENT THAT CAN BE EXPOSED TO NUMEROUS SURGERIES. DAMAGE FROM PROLONGED USE, MISUSE OR ROUGH HANDLING ARE LIKELY POTENTIAL FACTORS THAT COULD CONTRIBUTE TO THE REPORTED EVENT. WE RECOMMEND THAT ALL REUSABLE INSTRUMENTS BE ROUTINELY INSPECTED FOR WEAR AND DAMAGE AND REPLACED AS NECESSARY. BASED ON THIS INVESTIGATION, THE NEED FOR CORRECTIVE ACTION IS NOT INDICATED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE COMPLAINT WILL BE REOPENED. NO FURTHER INVESTIGATION IS WARRANTED FOR THIS COMPLAINT; HOWEVER, WE WILL CONTINUE TO MONITOR FUTURE COMPLAINTS AND INVESTIGATE AS NECESSARY. WE CONSIDER THIS INVESTIGATION CLOSED.

Description of Event or Problem · 0

IT WAS REPORTED THAT, DURING A TKA SURGERY, A GAP SPACER BROKE IN HALF. ALL PIECES WERE REMOVED WITH A TONSIL FROM PATIENT. THE SURGERY WAS RESUMED, WITHOUT ANY DELAY, USING A SMITH AND NEPHEW BACK-UP DEVICE. PATIENT WAS NOT INJURED BEYOND THE REPORTED PROBLEM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
270926 GAP SPACER TEMPLATE HWT SMITH & NEPHEW, INC. 71935180 21BDN2160 00885556578445

Patients

Seq Age Sex Outcome Treatment
1 Unknown Required Intervention