ELECTROSURGICAL GENERATOR "ESG-400"
Report
- Report Number
- 9610773-2022-00406
- Event Type
- Malfunction
- Date Received
- September 15, 2022
- Date of Event
- September 5, 2022
- Report Date
- September 15, 2022
- Manufacturer
- OLYMPUS WINTER & IBE GMBH
- Product Code
- GEI
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER
Narratives
THE SUSPECT MEDICAL DEVICE WAS NOT RETURNED TO THE MANUFACTURER FOR EVALUATION/INVESTIGATION BUT TO AN OLYMPUS REGIONAL REPAIR CENTER (RRC) IN CHINA (RETURNED TO RRC ON 2022-09-08). THE EVALUATION CONFIRMED THE OCCURRENCE OF ERROR MESSAGE E433 WHICH IS TRIGGERED BY THE GENERATOR¿S SAFETY SYSTEM. IN CASE OF CRITICAL ERRORS, THE SAFETY SYSTEM WILL NOT PERMIT ANY FURTHER USE OF THE GENERATOR UNTIL THE ERROR IS RECTIFIED. IN THE CASE AT HAND, THIS WAS CAUSED BY A DEFECTIVE GENERATOR BOARD. THUS, THIS EVENT/INCIDENT WAS ATTRIBUTED TO COMPONENT FAILURE. A MANUFACTURING AND QUALITY CONTROL REVIEW WAS PERFORMED FOR THE AFFECTED SERIAL NUMBER OF THE HF GENERATOR WITHOUT SHOWING ANY ABNORMALITIES. THE CASE WILL BE CLOSED ON OLYMPUS SIDE WITH NO FURTHER ACTIONS, BUT THE REPORTED EVENT/INCIDENT WILL BE RECORDED FOR TRENDING AND SURVEILLANCE PURPOSES. FURTHERMORE, THE USER WILL BE INFORMED ABOUT THE INVESTIGATION RESULTS. PLEASE NOTE: THIS REPORT IS BEING SUBMITTED ALTHOUGH THE SUSPECT MEDICAL DEVICE IS NOT MARKETED IN THE USA. HOWEVER, A SIMILAR DEVICE IS MARKETED. MODEL # / CATALOG #: WB91051W; BRAND NAME: ELECTROSURGICAL GENERATOR "ESG-400"; COMMON DEVICE NAME: HF-GENERATORS; 510(K): K203682; PRODUCT CODE: GEI.
OLYMPUS WAS INFORMED THAT DURING INSPECTION PRIOR TO USE FOR AN UNSPECIFIED PROCEDURE, THE ELECTROSURGICAL GENERATOR ¿ESG-400¿ DISPLAYED ERROR MESSAGE E433. THE INTENDED PROCEDURE WAS THEN SUCCESSFULLY COMPLETED WITH A SIMILAR DEVICE AND THERE WAS NO REPORT ABOUT AN ADVERSE EVENT OR PATIENT INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 270877 | ELECTROSURGICAL GENERATOR "ESG-400" | HF-GENERATORS | GEI | OLYMPUS WINTER & IBE GMBH | WB91051C |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown | OLYMPUS ULTRASONIC GENERATOR USG-400 |