FDA Adverse Event Other Summary report: N

TRACH CARE CLOSED SUCTION SYSTEM FOR ADULTS

MDR report key: 1541905 · Received November 10, 2009

Report

Report Number
8030647-2009-00015
Event Type
Other
Date Received
November 10, 2009
Date of Event
August 17, 2009
Report Date
October 12, 2009
Manufacturer
AVENT S. DE R.L. DE C.V.
Product Code
BSY
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
OK, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE USER HOSPITAL HAS REPORTED THIS INCIDENT AS "USER ERROR." A SAMPLE WILL NOT BE RETURNED FOR EVALUATION, AND THE LOT # OF THE DEVICE WAS NOT PROVIDED; THEREFORE, THE DEVICE HISTORY RECORD COULD NOT BE REVIEWED. THE DIRECTIONS FOR USE CONTAINS A WARNING THAT STATES "CAP ON TRACH CARE* T-PIECE PREVENTS CONTINUOUS FLOW THERAPY. REMOVE CAP BEFORE STARTING CONTINUOUS FLOW THERAPY. FAILURE TO REMOVE CAP PRIOR TO CONTINUOUS FLOW THERAPY MAY RESULT IN SERIOUS INJURY OR DEATH." WE WILL CONTINUE TO CLOSELY MONITOR THE FIELD PERFORMANCE OF THIS PRODUCT TO IDENTIFY EMERGING TRENDS AND IF APPLICABLE, ADDITIONAL INVESTIGATIONS AND CORRECTIVE ACTIONS WILL BE TAKEN.

Description of Event or Problem · 1

KIMBERLY CLARK RECEIVED A COMPLAINT INDICATING "TRACH COLLAR CHANGED TO IN-LINE SUCTION. EXHALATION CAP WAS LEFT ON, NOT ALLOWING FOR EXHALATION." THE PATIENT WAS BRADYCARDIC, UNRESPONSIVE, AND CYANOTIC. THE USER HOSPITAL REPORTED THIS INCIDENT AS A "USER ERROR" AS THE THERAPIST FAILED TO REMOVE THE CAP. THE PATIENT IMMEDIATELY RESPONDED WELL TO BAGGING WITH IN-LINE SUCTION REMOVED. NO OTHER PROBLEMS WERE NOTED. KIMBERLY-CLARK HAS NO INDEPENDENT KNOWLEDGE OF THE EVENT REPORTED BUT IS RELAYING THE INFORMATION THAT WAS PROVIDED BY THE USER FACILITY WHERE THE INCIDENT OCCURRED. THIS PRODUCT INCIDENT IS DOCUMENTED IN THE KIMBERLY-CLARK COMPLAINT DATABASE AND IDENTIFIED AS PIR (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TRACH CARE CLOSED SUCTION SYSTEM FOR ADULTS CLOSED SUCTION SYSTEM 14F - TPIECE BSY AVENT S. DE R.L. DE C.V.

Patients

Seq Age Sex Outcome Treatment
1 30 YR Required Intervention