FDA Adverse Event
Malfunction
Summary report: N
T:SLIM X2 INSULIN PUMP WITH CONTROL-IQ TECHNOLOGY
MDR report key: 15417869
·
Received September 14, 2022
Report
- Report Number
- 3013756811-2022-99787
- Event Type
- Malfunction
- Date Received
- September 14, 2022
- Date of Event
- August 21, 2022
- Report Date
- August 22, 2022
- Manufacturer
- TANDEM DIABETES CARE
- Product Code
- QFG
- UDI-DI
- 00850006613793
- PMA / PMN Number
- K201214
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 0
NO PRODUCT WAS RETURNED FOR EVALUATION. SHOULD NEW RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
Description of Event or Problem · 0
IT WAS REPORTED THAT A MINIMUM FILL NOTIFICATION OCCURRED AFTER THE USER FILLED THE CARTRIDGE WITH 250 UNITS OF INSULIN DURING THE LOAD SEQUENCE. A NEW CARTRIDGE WAS LOADED TO RESOLVE THE ISSUE. CUSTOMER¿S BLOOD GLUCOSE LEVEL RANGED FROM 200-291 MG/DL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2643167 | T:SLIM X2 INSULIN PUMP WITH CONTROL-IQ TECHNOLOGY | ALTERNATE CONTROLLER ENABLED INFUSION PUMP | QFG | TANDEM DIABETES CARE | 1000354 | W0469580 | 00850006613793 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 7 YR | Female |