I-STAT CHEM8+ CARTRIDGE
Report
- Report Number
- 2245578-2022-00134
- Event Type
- Malfunction
- Date Received
- September 14, 2022
- Date of Event
- September 5, 2022
- Report Date
- September 20, 2022
- Manufacturer
- ABBOTT POINT OF CARE
- Product Code
- JGS
- UDI-DI
- 10054749001917
- PMA / PMN Number
- K053110
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
APOC INCIDENT: # (B)(4). THE INVESTIGATION WAS COMPLETED ON 19-SEP-2022. A REVIEW OF THE DEVICE HISTORY RECORD CONFIRMED THE LOT PASSED FINISHED GOODS RELEASE CRITERIA. RETAINED CARTRIDGE TESTING MET THE ACCEPTANCE CRITERIA FOUND IN Q04.01.003 REV. AK, APPENDIX 1- PRODUCT COMPLAINT LEVEL 2 AND LEVEL 3 INVESTIGATION PROCEDURE. NO DEFICIENCY HAS BEEN IDENTIFIED.
APOC INCIDENT # (B)(4). APOC LABELING WILL BE EVALUATED DURING THE INVESTIGATION AS PERTAINING TO THE EVENT.
ON (B)(6) 2022, ABBOTT POINT OF CARE WAS CONTACTED BY A CUSTOMER REGARDING I-STAT CHEM8+ CARTRIDGES THAT YIELDED SUSPECTED DISCREPANT RESULTS ON A PATIENT. THERE WAS NO PATIENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT. I-STAT 7.5 A, CRRT FILTERED 4.1 NI, I-STAT 4.1 B. THERE ARE NO INJURIES ASSOCIATED WITH THIS EVENT. AT THIS TIME THERE IS NO REASON TO SUSPECT A MALFUNCTION EXISTS. THE REPORTING DECISION WAS BASED ON LIMITED INFORMATION AVAILABLE THAT SUGGESTS THE PRODUCT WAS NOT PERFORMING WITHIN THE VARIABILITY OF THE ASSAY. THE INVESTIGATION IS UNDERWAY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2643155 | I-STAT CHEM8+ CARTRIDGE | CHEM8+ CARTRIDGE | JGS | ABBOTT POINT OF CARE | NA | H22021A | 10054749001917 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |