- Report Number
- 9610877-2022-57793
- Event Type
- Malfunction
- Date Received
- September 14, 2022
- Date of Event
- August 26, 2022
- Report Date
- September 14, 2022
- Manufacturer
- HOYA CORPORATION PENTAX TOKYO OFFICE
- Product Code
- EOB
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MD, US
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
THIS DEVICE IS CLASSIFIED AS IMPORT FOR EXPORT, THEREFORE 510K IS NOT APPLICABLE. MODEL VNL-1570STK IS AVAILABLE IN THE USA WITH A 510K NUMBER K162151. WE CHECKED THE RETURNED UNIT AND CONFIRMED THAT THE CCD DRIVER PCB AS DEFECTIVE. BASED ON THE RESULT, WE CONCLUDED THAT IT WAS CAUSED DUE TO THE EXCESSIVE FORCE APPLIED ON THE CCD DRIVER PCB. IN ADDITION, WE CONFIRMED THAT THE INSERTION FLEXIBLE TUBE (IFT) CRUSHED, THE SEGMENT CRUSHED, AND THE LIGHT GUIDE CABLE FOR PRONG LEAK; HOWEVER, THEY ARE NOT THE MAIN CAUSE, AND/OR IRRELEVANT TO THE ALLEGED COMPLAINT. BASED ON THE TECHNICAL REPORT ""HR-RPT-0586(IMAGE FAILURE)"" AND/OR THE RISK ANALYSIS RESULTS, IT WAS EVALUATED TO SUBMIT MDR.
THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2364642 | PENTAX | VIDEO NASO PHARYNGO LARYNGOSTROBOSCOPE | EOB | HOYA CORPORATION PENTAX TOKYO OFFICE | VNL-1070STK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |