FDA Adverse Event Malfunction Summary report: N

BD LUER-LOK¿ SYRINGE

MDR report key: 15407611 · Received September 12, 2022

Report

Report Number
1213809-2022-00564
Event Type
Malfunction
Date Received
September 12, 2022
Date of Event
August 9, 2022
Report Date
September 14, 2022
Manufacturer
BECTON DICKINSON MEDICAL SYSTEMS
Product Code
FMF
UDI-DI
00382903029952
PMA / PMN Number
K172670
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H6: INVESTIGATION SUMMARY ELEVEN SAMPLES IN SEALED PACKAGES WERE PROVIDED TO OUR QUALITY TEAM FOR INVESTIGATION. THROUGH VISUAL INSPECTION, IT WAS OBSERVED THAT EIGHT SAMPLES HAD A SKEWED SCALE MARKING. POTENTIAL ROOT CAUSE FOR THE SKEWED DEFECT IS ASSOCIATED WITH THE PRINTER PROCESS. THESE CONDITIONS ARE OCCURRING BELOW THEIR EXPECTED FREQUENCY SO NO CORRECTIVE ACTION IS REQUIRED AT THIS TIME. BATCH 2124536 IS CONSIDERED IN COMPLIANCE WITH OUR PRODUCT SPECIFICATION REQUIREMENTS. A DEVICE HISTORY RECORD REVIEW WAS COMPLETED FOR PROVIDED LOT NUMBER 2124536. A REVIEW SHOWED NO REJECTED INSPECTIONS OR QUALITY ISSUES DURING THE PRODUCTION THAT COULD HAVE CONTRIBUTED TO THE REPORTED DEFECT.

Additional Manufacturer Narrative · 0

A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED, BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE BD LUER-LOK¿ SYRINGE HAVE A MISPRINT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: WE HAVE FOUND MULTIPLES OF THIS PRODUCT THAT HAVE A MISPRINT OF THE GRADUATION LINES AND NUMBERS.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE BD LUER-LOK¿ SYRINGE HAVE A MISPRINT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: WE HAVE FOUND MULTIPLES OF THIS PRODUCT THAT HAVE A MISPRINT OF THE GRADUATION LINES AND NUMBERS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2533567 BD LUER-LOK¿ SYRINGE PISTON SYRINGE FMF BECTON DICKINSON MEDICAL SYSTEMS 2124536 00382903029952

Patients

Seq Age Sex Outcome Treatment
1 Unknown