FDA Adverse Event Malfunction Summary report: N

BD LUER-LOCK¿ FEMALE ADAPTER

MDR report key: 15407166 · Received September 12, 2022

Report

Report Number
2243072-2022-01490
Event Type
Malfunction
Date Received
September 12, 2022
Date of Event
August 18, 2022
Report Date
August 20, 2022
Manufacturer
BECTON DICKINSON
Product Code
FPA
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

OEM MANUFACTURER: THE MANUFACTURING LOCATION FOR THIS PRODUCT IS (B)(4). THIS SITE IS AN OEM MANUFACTURING SITE. THEREFORE, (B)(4). MEDICAL DEVICE EXPIRATION DATE: NA. INVESTIGATION SUMMARY: A COMPLAINT OF FOREIGN MATTER BEING FOUND INSIDE A LUER WAS RECEIVED FROM THE CUSTOMER. PHOTOS WERE PROVIDED FOR INVESTIGATION WHERE A CLEAR MATERIAL COULD BE SEEN INSIDE THE LUER HOLE. A DEVICE HISTORY REVIEW WAS PERFORMED FOR LOT#: 2041965 AND NO QUALITY NOTIFICATIONS WERE OPENED AT NAMC RELATED TO THIS CONDITION. AN INVESTIGATION WAS PERFORMED THAT INDICATED THIS CONDITION COULD OCCUR DUE TO A PARTIALLY BROKEN PIN OR FLASH AROUND THE CORE DUE TO A BAD SHUT OFF GENERATED BY A DAMAGE PIN. SINCE NO PHYSICAL SAMPLES WERE AVAILABLE FOR REVIEW AT NAMC, AND NO CAVITY NUMBERS WERE PROVIDED, THE DEFECT COULD NOT BE CONFIRMED. ADDITIONALLY, IT IS UNLIKELY TO BE A TOOLING RELATED FLASH DUE TO THE GEOMETRY DISPLAYED ON THE DEFECTIVE IMAGE. FURTHERMORE, FROM THE MOLD PREVENTATIVE MAINTENANCE RECORD, THERE IS NO INDICATION OF THIS CONDITION BEING PRESENT AT NAMC. THIS INCIDENT HAS BEEN ADDED TO OUR DATABASE OF REPORTED INCIDENTS. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.

Description of Event or Problem · 0

IT WAS REPORTED THAT FOREIGN MATTER WAS FOUND IN THE LUER OF THE BD LUER-LOCK¿ FEMALE ADAPTER. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "DEFECT - DEBRIS IN LUER THRU HOLE"

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2591690 BD LUER-LOCK¿ FEMALE ADAPTER INTRAVASCULAR ADMINISTRATION SET FPA BECTON DICKINSON 2041965

Patients

Seq Age Sex Outcome Treatment
1 Unknown