NUVASIVE MODULUS XLIF INTERBODY SYSTEM
Report
- Report Number
- 2031966-2022-00177
- Event Type
- Malfunction
- Date Received
- September 9, 2022
- Date of Event
- August 12, 2022
- Report Date
- September 9, 2022
- Manufacturer
- NUVASIVE, INCORPORATED
- Product Code
- MAX
- UDI-DI
- 00887517452764
- PMA / PMN Number
- K201692
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
THE DEVICE WAS RETURNED AND THE COMPLAINT WAS CONFIRMED. DHR REVIEW OF THE REPORTED DEVICE IDENTIFIED WITH RELEASE TO THE FIELD IN 2017. EXAMINATION OF THE RETURNED DEVICE WAS COMPLETED BY NUVASIVE AND THE FRACTURE PATTERN IDENTIFIED IS CONSISTENT WITH EXCESSIVE FORCE. REVIEW OF THE RETURNED DEVICE AND REPORTED EVENT SUGGESTS THE FRACTURE WAS THE RESULT OF OFF ANGLED EXCESSIVE FORCE AND CONSIDERED AN INADVERTENT USE ERROR. NO ADDITIONAL INVESTIGATION NECESSARY. LABELING REVIEW: "PATIENT EDUCATION: PREOPERATIVE INSTRUCTIONS TO THE PATIENT ARE ESSENTIAL. THE PATIENT SHOULD BE MADE AWARE OF THE POTENTIAL RISKS OF THE SURGERY." "DO NOT IMPLANT THE INSTRUMENTS: COMPLICATIONS TO THE PATIENT MAY INCLUDE, BUT ARE NOT LIMITED TO: BREAKAGE OF THE DEVICE, WHICH COULD MAKE NECESSARY REMOVAL DIFFICULT OR SOMETIMES IMPOSSIBLE, WITH POSSIBLE CONSEQUENCES OF LATE INFECTION AND MIGRATION. BREAKAGE COULD CAUSE INJURY TO THE PATIENT." "PRE-OPERATIVE WARNINGS: THE METHOD OF USE FOR THE INSTRUMENTS ARE TO BE DETERMINED BY THE USER¿S EXPERIENCE AND TRAINING IN SURGICAL PROCEDURES. THE INSTRUMENTS SHOULD BE TREATED AS ANY PRECISION INSTRUMENT AND SHOULD BE CAREFULLY PLACED ON TRAYS, CLEANED AFTER EACH USE, AND STORED, ACCORDING TO GENERALLY ACCEPTED HOSPITAL METHODS AND PRACTICES. THE INSTRUMENTS SHOULD BE CAREFULLY EXAMINED PRIOR TO USE FOR FUNCTIONALITY, EXCESSIVE WEAR, OR DAMAGE. A DAMAGED INSTRUMENT SHOULD NOT BE USED AS THIS MAY INCREASE THE RISK OF MALFUNCTION AND POTENTIAL PATIENT INJURY. CARE SHOULD BE USED DURING SURGICAL PROCEDURES TO PREVENT DAMAGE TO THE DEVICES AND INJURY TO THE PATIENT." "INTRA-OPERATIVE WARNINGS: THE PHYSICIAN SHOULD TAKE PRECAUTIONS AGAINST PUTTING UNDUE STRESS ON THE SPINAL AREA WITH INSTRUMENTS. ANY SURGICAL TECHNIQUE SHOULD BE CAREFULLY FOLLOWED. IT IS IMPORTANT THAT THE SURGEON EXERCISE EXTREME CAUTION WHEN WORKING IN CLOSE PROXIMITY TO VITAL ORGANS, NERVES, OR VESSELS, AND THAT THE FORCE APPLIED TO THE INSTRUMENTATION IS NOT EXCESSIVE, TO PREVENT POTENTIAL INJURY TO THE PATIENT. OVER-BENDING, NOTCHING, STRIKING, AND/OR SCRATCHING OF IMPLANTS WITH ANY INSTRUMENT SHOULD BE AVOIDED TO REDUCE THE RISK OF BREAKAGE. THE PHYSICAL CHARACTERISTICS REQUIRED FOR MANY INSTRUMENTS DO NOT PERMIT THEM TO BE MANUFACTURED FROM IMPLANTABLE MATERIALS. IF ANY BROKEN FRAGMENTS OF INSTRUMENTS REMAIN IN THE BODY OF A PATIENT, THEY COULD CAUSE ALLERGIC REACTIONS OR INFECTIONS. IF AN INSTRUMENT BREAKS IN SURGERY AND FRAGMENTS GO INTO THE PATIENT, THESE PIECES SHOULD BE REMOVED PRIOR TO CLOSURE AND SHOULD NOT BE IMPLANTED."
ON (B)(6) 2022, DURING A SPINAL PROCEDURE THE TOP PITUITARY TIP BROKE OFF. THE PIECES WERE RETRIEVED AND NO PATIENT INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1138954 | NUVASIVE MODULUS XLIF INTERBODY SYSTEM | INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT, LUMBAR | MAX | NUVASIVE, INCORPORATED | 6940429 | FD5975440 | 00887517452764 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Prefer Not To Disclose |