BD VACUTAINER® SST¿ II ADVANCE PLUS BLOOD COLLECTION TUBES
Report
- Report Number
- 9617032-2022-00842
- Event Type
- Malfunction
- Date Received
- September 9, 2022
- Date of Event
- August 15, 2022
- Report Date
- August 22, 2022
- Manufacturer
- BECTON, DICKINSON AND COMPANY (BD)
- Product Code
- JKA
- UDI-DI
- 50382903679534
- PMA / PMN Number
- BK050036
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TW
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
THERE WERE MULTIPLE LOT NUMBERS REPORTED TO BE INVOLVED. THE INFORMATION FOR EACH LOT NUMBER IS AS FOLLOWS: MEDICAL DEVICE LOT #: 2080645. MEDICAL DEVICE EXPIRATION DATE: 2023-09-30. DEVICE MANUFACTURE DATE: 2022-03-21. MEDICAL DEVICE LOT #: 2020073. MEDICAL DEVICE EXPIRATION DATE: 2023-07-31. DEVICE MANUFACTURE DATE: 2022-01-20. INVESTIGATION SUMMARY: MATERIAL NO. 367953. BATCH NO. 2080645, 2020073. BD HAD NOT RECEIVED SAMPLES, BUT THREE (3) PHOTOS WERE PROVIDED BY YOUR FACILITY FOR INVESTIGATION. THE PHOTOS WERE EVALUATED AND THE INDICATED FAILURE MODES FOR COLLAPSED TUBES AND UNDERFILL WITH THE INCIDENT LOTS WAS OBSERVED. RETENTION SAMPLES WERE SELECTED FROM BD INVENTORY FOR VISUAL EXAMINATION AND UPON COMPLETION, COLLAPSED TUBES WERE NOT OBSERVED. ADDITIONALLY, TWENTY (20) RETENTION SAMPLES OF EACH LOT FROM BD INVENTORY WERE EVALUATED BY FUNCTIONAL TESTING AND THE INDICATED FAILURE MODE OF UNDERFILL WAS NOT OBSERVED AS ALL SAMPLES MET SPECIFICATIONS. A REVIEW OF THE MANUFACTURING RECORDS WAS COMPLETED FOR THE INCIDENT LOT AND NO ISSUES WERE IDENTIFIED. BD IS CONTINUALLY MONITORING COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION IN ORDER TO IDENTIFY EMERGING TRENDS. NOTE THAT THERE ARE POTENTIALLY VARYING DEGREES OF TUBE COLLAPSE, BASED ON THE TEMPERATURE AND DURATION OF TIME AT ELEVATED TEMPERATURE. MILDLY COLLAPSED TUBES WILL RETAIN VACUUM AND ONLY SHOW MINIMAL DEFORMATION. MILDLY COLLAPSED TUBES WILL DRAW APPROPRIATELY BUT MAY NOT BE ABLE TO BE PROCESSED ON LABORATORY INSTRUMENTS (E.G. MAY BE SLIGHTLY BOWED AND NO LONGER FIT IN A RACK). AS LONG AS THE VACUUM IS RETAINED AND THE TUBE FILLS APPROPRIATELY, THE TUBE WILL FUNCTION PROPERLY. FULLY COLLAPSED TUBES APPEAR FLATTENED, TWISTED, AND VISIBLY DEFORMED. FULLY COLLAPSED TUBES RETAIN LITTLE OR NO VACUUM AND THEREFORE CANNOT BE DRAWN TO AN APPRECIABLE VOLUME. THESE TUBES ARE EASILY IDENTIFIED BEFORE USE BY THE HEALTHCARE WORKER, SO THERE IS NO POTENTIAL IMPACT TO THE PATIENT. THIS COMPLAINT HAS BEEN CONFIRMED FOR THE INDICATED FAILURE MODES OF UNDERFILL AND COLLAPSED TUBES FROM RETURNED PHOTOS. BD WAS NOT ABLE TO IDENTIFY A ROOT CAUSE FOR UNDERFILL. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.
IT WAS REPORTED WHEN USING THE BD VACUTAINER® SST¿ II ADVANCE PLUS BLOOD COLLECTION TUBES, THE DEVICE EXPERIENCED UNDERFILL OR LOW DRAW OF A TUBE, AS WELL AS COLLAPSED TUBES. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER. THE CUSTOMER STATED: BATCH 2080645: DEFORMATION AND INSUFFICIENT VACUUM. BATCH 2020073: INSUFFICIENT VACUUM.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1806747 | BD VACUTAINER® SST¿ II ADVANCE PLUS BLOOD COLLECTION TUBES | BLOOD SPECIMEN COLLECTION DEVICE | JKA | BECTON, DICKINSON AND COMPANY (BD) | 367953 | 2020073 | 50382903679534 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |