CADD LEGACY PLUS PUMP
Report
- Report Number
- 3012307300-2022-18242
- Event Type
- Malfunction
- Date Received
- September 8, 2022
- Report Date
- October 19, 2022
- Manufacturer
- ST PAUL
- Product Code
- FRN
- PMA / PMN Number
- K982838
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
UNIQUE IDENTIFIER (UDI) IS UNKNOWN. NO INFORMATION HAS BEEN PROVIDED TO DATE. INVESTIGATION INCLUDING ROOT CAUSE ANALYSIS IS IN PROGRESS. A SUPPLEMENTAL MDR WILL BE FILED AS NECESSARY IN ACCORDANCE WITH 21 CFR 803.56 WHEN ADDITIONAL REPORTABLE INFORMATION BECOMES AVAILABLE.
OTHER TEXT: D10: DEVICE AVAILABLE FOR EVALUATION: UPDATED. H3: DEVICE EVALUATED BY MANUFACTURER: UPDATED. H6: EVENT PROBLEM AND EVALUATION CODES: UPDATED. H10: DEVICE EVALUATION: THE DEVICE WAS RETURNED TO MANUFACTURING FOR INVESTIGATION. VISUAL INSPECTION AND FUNCTIONAL TESTING WERE PERFORMED. VISUAL INSPECTION OF THE RETURNED DEVICE CONFIRMED THAT THE TAMPER SEALS AND PHYSICAL CONDITION WERE ALL INTACT UPON RECEIPT. ERROR CODES WERE FOUND IN THE EHL (EVENT HISTORY LOG) MULTIPLE TIMES. THE CUSTOMER REPORTED PROBLEM WAS DUPLICATED DURING THE INVESTIGATION AS THE OCCLUSION DETECTION LEVEL OF THE DOWNSTREAM OCCLUSION SENSOR (DSO) WAS LOW LEVEL AND WITHOUT THE STANDARD. THE ROOT CAUSE OF THE REPORTED EVENT IS UNKNOWN. REPLACE THE DOWNSTREAM SENSOR. A DEVICE HISTORY RECORD DHR REVIEW WAS NOT CONDUCTED, BASED UPON REVIEW OF THE INFORMATION PROVIDED BY THE CUSTOMER, AS IT DOES NOT INDICATE A PROBLEM WITH THE INITIAL MANUFACTURE OR PRIOR REPAIR OF THE DEVICE., CORRECTED DATA: G5: PREMARKET 510K NUMBER: K982838.
IT WAS REPORTED THAT THE DEVICE HAD A HIGH PRESSURE ALARM. NO PATIENT INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2122350 | CADD LEGACY PLUS PUMP | PUMP, INFUSION | FRN | ST PAUL | 6500 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |