FDA Adverse Event Malfunction Summary report: N

CADD LEGACY PLUS PUMP

MDR report key: 15385406 · Received September 8, 2022

Report

Report Number
3012307300-2022-18242
Event Type
Malfunction
Date Received
September 8, 2022
Report Date
October 19, 2022
Manufacturer
ST PAUL
Product Code
FRN
PMA / PMN Number
K982838
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

UNIQUE IDENTIFIER (UDI) IS UNKNOWN. NO INFORMATION HAS BEEN PROVIDED TO DATE. INVESTIGATION INCLUDING ROOT CAUSE ANALYSIS IS IN PROGRESS. A SUPPLEMENTAL MDR WILL BE FILED AS NECESSARY IN ACCORDANCE WITH 21 CFR 803.56 WHEN ADDITIONAL REPORTABLE INFORMATION BECOMES AVAILABLE.

Additional Manufacturer Narrative · 0

OTHER TEXT: D10: DEVICE AVAILABLE FOR EVALUATION: UPDATED. H3: DEVICE EVALUATED BY MANUFACTURER: UPDATED. H6: EVENT PROBLEM AND EVALUATION CODES: UPDATED. H10: DEVICE EVALUATION: THE DEVICE WAS RETURNED TO MANUFACTURING FOR INVESTIGATION. VISUAL INSPECTION AND FUNCTIONAL TESTING WERE PERFORMED. VISUAL INSPECTION OF THE RETURNED DEVICE CONFIRMED THAT THE TAMPER SEALS AND PHYSICAL CONDITION WERE ALL INTACT UPON RECEIPT. ERROR CODES WERE FOUND IN THE EHL (EVENT HISTORY LOG) MULTIPLE TIMES. THE CUSTOMER REPORTED PROBLEM WAS DUPLICATED DURING THE INVESTIGATION AS THE OCCLUSION DETECTION LEVEL OF THE DOWNSTREAM OCCLUSION SENSOR (DSO) WAS LOW LEVEL AND WITHOUT THE STANDARD. THE ROOT CAUSE OF THE REPORTED EVENT IS UNKNOWN. REPLACE THE DOWNSTREAM SENSOR. A DEVICE HISTORY RECORD DHR REVIEW WAS NOT CONDUCTED, BASED UPON REVIEW OF THE INFORMATION PROVIDED BY THE CUSTOMER, AS IT DOES NOT INDICATE A PROBLEM WITH THE INITIAL MANUFACTURE OR PRIOR REPAIR OF THE DEVICE., CORRECTED DATA: G5: PREMARKET 510K NUMBER: K982838.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE DEVICE HAD A HIGH PRESSURE ALARM. NO PATIENT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2122350 CADD LEGACY PLUS PUMP PUMP, INFUSION FRN ST PAUL 6500

Patients

Seq Age Sex Outcome Treatment
1 Unknown