CHECK-IT
Report
- Report Number
- 3016521623-2022-00409
- Event Type
- Malfunction
- Date Received
- September 8, 2022
- Date of Event
- August 30, 2022
- Report Date
- September 7, 2022
- Manufacturer
- LUCIRA HEALTH, INC.
- Product Code
- QJR
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
PRODUCT HAS NOT BEEN RETURNED AND NO HARM(S) REPORTED. THEREFORE, A DHR REVIEW WAS COMPLETED FOR KIT LOT NUMBER K08A112707224M7 WITH 0 DISCREPANCIES FOUND (ALL LOTS PASSED IN TOTAL PER THE SAMPLING PLAN). A REVIEW OF EXISTING CAPAS, SCARS AND NCMRS WERE COMPLETED AND THERE ARE NO PRIOR RECORDS RELATED TO "FALSE POSITIVE" FAILURE MODE FOR THIS LOT. KIT LOT NUMBER K08A112707224M7, HAS 0 ADDITIONAL COMPLAINTS FROM THIS CUSTOMER NOR ANY OTHER CUSTOMER ASSOCIATED WITH A "FALSE POSITIVE" FAILURE MODE PRIOR TO THE REPORTED RECEIPT DATE OF AUGUST 30, 2022. BASED ON REVIEW OF THE PRODUCT'S FMEAS AND RISK ASSESSMENT DOCUMENTATION, FALSE POSITIVE TEST RESULTS ARE A KNOWN POSSIBLE OUTCOME REGARDING THIS ISSUE UNDER EVALUATION, REFER TO FMEA-001 AND FMEA-004. THE COMPLAINT RATE FOR "FALSE POSITIVE" IS UNDER THE EXPECTED THRESHOLD OF 2% (LABEL CLAIM)/1% (INTERNAL WARNING LIMIT). ACCORDING TO THE FAILURE RATE (0.08%) FOR THIS LOT, THE NUMBER OF FALSE POSITIVES (3-TOTAL FALSE POSITIVES REPORTED FROM SAME LOT) IN RELATION TO THE TOTAL LOT QUANTITY PRODUCED (3,990) IS UNDER THE EXPECTED THRESHOLD. DHR REVIEW KIT LOT# K08A112707224M7: SAMPLE VIAL LOT DHR REVIEW: 2112125, 2204175 (ASSOCIATED INTERNAL LOT # 210762-2KC, 220305-8D) TEST LOT DHRS REVIEW: 2203578 (ASSOCIATED INTERNAL LOT # 220305-27S) BASED ON THE LIMITED INFORMATION AVAILABLE, ROOT CAUSE CANNOT BE DETERMINED. POTENTIAL ROOT CAUSES INCLUDE BUT ARE NOT LIMITED TO: ENVIRONMENTAL CONTAMINATION, LOW VIRAL LOAD, AND/OR DEVICE FAILURE. LUCIRA HEALTH WILL CONTINUE TO MONITOR TRENDS RELATED TO FALSE POSITIVE RESULTS IN ACCORDANCE WITH POST-MARKET SURVEILLANCE PROCESS. A SUPPLEMENTAL REPORT WILL BE FILED IF ANY FURTHER INVESTIGATION AND/OR ADDITIONAL INFORMATION IS OBTAINED. BASED ON THE COMPLAINT REVIEW, NO HARM WAS REPORTED WITHIN THE COMPLAINT. THIS DEVICE IS MARKETED UNDER EUA 210196 CHECK-IT. D4-EXPIRATION DATE OF KIT LOT NUMBER K08A112707224M7-03/18/2023.
ONE DEVICE REPORTED AS HAVING AN ALLEGED FALSE POSITIVE RESULT. THE COMPLAINTANT REPORTED THAT AN INDIVIDUAL PERFORMED A LUCIRA TEST (KIT LOT NUMBER K08A112707224M7) WITH A POSTITIVE RESULT. THE CUSTOMER RETESTED WITH AN ACCULA (PCR) TEST WITH A NEGATIVE RESULT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2237436 | CHECK-IT | COVID-19 TEST KIT | QJR | LUCIRA HEALTH, INC. | K08A112707224M7 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown | Other |