FDA Adverse Event Malfunction Summary report: N

I-STAT CHEM8+ CARTRIDGE

MDR report key: 15382987 · Received September 8, 2022

Report

Report Number
2245578-2022-00130
Event Type
Malfunction
Date Received
September 8, 2022
Date of Event
September 3, 2022
Report Date
November 18, 2022
Manufacturer
ABBOTT POINT OF CARE
Product Code
JGS
UDI-DI
10054749005519
PMA / PMN Number
K183688
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

APOC INCIDENT: # (B)(4). ANALYZER INVESTIGATION (896411): THE INVESTIGATION WAS COMPLETED ON 16-NOV-2022. THE CUSTOMER REPORTED THAT ANALYZER S/N (B)(6) WAS GIVING DISCREPANT PATIENT RESULTS FOR POTASSIUM (K) AND CREATININE (CREA) WHILE RUNNING I-STAT CHEM8+ CARTRIDGES FROM LOTS H22138 AND H22164 WITH I-STAT CONTROL LEVEL 1 (LOT 101144), 3 (LOT 121145), AND CALIBRATION VERIFICATION SET (LOT 21258). ADDITIONALLY, THE ANALYZER PRODUCED HIGH CHLORIDE (CI) AND LOW UREA NITROGEN (BUN) RESULTS. FAILURE ANALYSIS COULD NOT BE PERFORMED AS ANALYZER S/N (B)(6) HAS NOT BEEN RETURNED TO FLEX. A ROCKETWARE SEARCH SPANNING THREE MONTHS REVEALED NO SIMILAR INCIDENTS AND NO EVIDENCE OF A TREND. NO DEFICIENCY HAS BEEN IDENTIFIED.

Additional Manufacturer Narrative · 0

APOC INCIDENT: # (B)(4). THE INVESTIGATION WAS COMPLETED ON 28-SEP-2022. A REVIEW OF THE DEVICE HISTORY RECORD CONFIRMED THE CARTRIDGE LOT PASSED FINISHED GOODS RELEASE CRITERIA. RETAINED CARTRIDGE TESTING MET THE ACCEPTANCE CRITERIA FOUND IN Q04.01.003 REV. AK, APPENDIX 1- PRODUCT COMPLAINT LEVEL 2 AND LEVEL 3 INVESTIGATION PROCEDURE. NO DEFICIENCY HAS BEEN IDENTIFIED.

Additional Manufacturer Narrative · 0

APOC INCIDENT # (B)(4). APOC LABELING WILL BE EVALUATED DURING THE INVESTIGATION AS PERTAINING TO THE EVENT. I-STAT PRODUCT INFORMATION: PRODUCT# 04P75-35, SERIAL # (B)(4). 26-SEP-2011. APOC INCIDENT # (B)(4).

Description of Event or Problem · 0

ON 06-SEP-2022, ABBOTT POINT OF CARE WAS CONTACTED BY A CUSTOMER REGARDING I-STAT CHEM8+ CARTRIDGES AND I-STAT SN (B)(4) THAT YIELDED A SUSPECTED DISCREPANT RESULT OF 2.5 ON A PATIENT. THERE WAS NO PATIENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT. RETURN PRODUCT IS AVAILABLE FOR INVESTIGATION. ANALYZER SN (B)(4) IS BEING REPLACED AT NO CHARGE AND WILL BE INVESTIGATED SEPARATELY. (B)(6). THERE ARE NO INJURIES ASSOCIATED WITH THIS EVENT. AT THIS TIME THERE IS NO REASON TO SUSPECT A MALFUNCTION EXISTS. THE REPORTING DECISION WAS BASED ON LIMITED INFORMATION AVAILABLE THAT SUGGESTS THE PRODUCT WAS NOT PERFORMING WITHIN THE VARIABILITY OF THE ASSAY. THE INVESTIGATION IS UNDERWAY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
280807 I-STAT CHEM8+ CARTRIDGE CHEM8+ CARTRIDGE JGS ABBOTT POINT OF CARE NA H22138 10054749005519

Patients

Seq Age Sex Outcome Treatment
1 83 YR Female