T:SLIM X2 INSULIN PUMP WITH CONTROL-IQ TECHNOLOGY
Report
- Report Number
- 3013756811-2022-99419
- Event Type
- Malfunction
- Date Received
- September 8, 2022
- Date of Event
- August 15, 2022
- Report Date
- August 19, 2022
- Manufacturer
- TANDEM DIABETES CARE
- Product Code
- QFG
- UDI-DI
- 00850006613793
- PMA / PMN Number
- K201214
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
NO PRODUCT WAS RETURNED FOR EVALUATION. SHOULD NEW RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PER TANDEM USER GUIDE: ONLY USE U-100 HUMALOG OR U-100 NOVOLOG WITH YOUR PUMP. ONLY U-100 HUMALOG AND NOVOLOG HAVE BEEN TESTED AND FOUND TO BE COMPATIBLE FOR USE IN THE PUMP. USE OF INSULIN WITH LESSER OR GREATER CONCENTRATION CAN RESULT IN UNDER DELIVERY OR OVER DELIVERY OF INSULIN. THIS CAN CAUSE HYPOGLYCEMIA (LOW BG) OR HYPERGLYCEMIA (HIGH BG) EVENTS.
IT WAS REPORTED THAT INTERMITTENT OCCLUSION ALARMS OCCURRED. CUSTOMER RAISED THE TEMP RATE AND CHANGED THE PUMP SUPPLIES TO ADDRESS THE ISSUE. REPORTEDLY, THE CUSTOMER IS USING FIASP INSULIN. TANDEM CLINICAL SUPPORT INFORMED CUSTOMER THAT FIASP IS OFF LABEL PER THE USER GUIDE. CUSTOMER¿S BLOOD GLUCOSE LEVEL WAS 110-159 MG/DL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2107851 | T:SLIM X2 INSULIN PUMP WITH CONTROL-IQ TECHNOLOGY | ALTERNATE CONTROLLER ENABLED INFUSION PUMP | QFG | TANDEM DIABETES CARE | 1000354 | 00850006613793 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 57 YR | Male | INFUSION SET: TRUSTEELINSULIN: FIASP |