EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE
Report
- Report Number
- 9610595-2022-01626
- Event Type
- Malfunction
- Date Received
- September 7, 2022
- Report Date
- February 15, 2023
- Manufacturer
- AIZU OLYMPUS CO., LTD.
- Product Code
- FDS
- UDI-DI
- 04953170305290
- PMA / PMN Number
- K131780
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
INDIANA UNIVERSITY HEALTH (IU HEALTH) BLOOMINGTON ENDOSCOPY CENTER (BEC) HAD CONCERNS CONNECTED WITH REPROCESSING OF DEVICES AT THE FACILITY. ARISING FROM THIS CONCERN, DEVICES/EVENTS HAVE BEEN IDENTIFIED FOR THE FACILITY AT THIS TIME AND ARE CAPTURED IN MEDWATCHES AS BELOW: (B)(6) CF-HQ190L (B)(4), (B)(6) PCF-H190DL (B)(4), (B)(6) PCF-H190DL (B)(4), (B)(6) PCF-H190DL (B)(4), (B)(6) PCF-H190DL (B)(4), (B)(6) GIF-H190 (B)(4). C22249083 GIF-H190 2742144 (B)(6), GIF-H190 2163363 (B)(6), GIF-H190 2742223, (B)(6) GIF-XP190N 2722764 (B)(6), GIF-H190 2742228, (B)(6) GIF-H190 2742219 THIS MEDWATCH IS FOR PATIENT IDENTIFIER: C22249098. THIS DEVICE IS OWNED BY INDIANA UNIVERSITY HEALTH BLOOMINGTON INC.; HOWEVER, IT IS PHYSICALLY LOCATED AT BEC AND BEING USED AT THAT FACILITY. THE DEVICE IS NOT RETURNED, AS SUCH A DEFINITIVE ROOT CAUSE OF THE REPORTED COMPLAINT CANNOT BE DETERMINED AT THIS TIME. THIS EVENT IS UNDER INVESTIGATION. A SUPPLEMENTAL REPORT WILL BE SUBMITTED UPON COMPLETION OF THE INVESTIGATION OR UPON RECEIVING ADDITIONAL INFORMATION. IN-SERVICE WAS PERFORMED BY THE OLYMPUS ENDOSCOPY SUPPORT SPECIALIST (ESS) FOR THE FACILITY¿S REPROCESSING. THE CLEANING, DISINFECTION, AND STERILIZATION INFORMATION WAS REVIEWED DURING IN-SERVICE AND DOCUMENTED. BELOW ARE THE OBSERVATIONS MADE: DURING REPROCESSING IN-SERVICE IT WAS OBSERVED THAT THE CUSTOMER KEEPS THE AUXILIARY WATER TUBE ON DURING LEAK TESTING. THE CUSTOMER DOES NOT USE LINT FREE CLOTHS DURING THE MANUAL CLEANING PROCESS. THE CUSTOMER USES THE CHANNEL OPENING CLEANING BRUSH PRIOR TO USING THE CHANNEL CLEANING BRUSH. THE BRUSHING IS DONE IN THE INCORRECT ORDER. THE CUSTOMER USES OLYMPUS BRUSHES. THE CUSTOMER ALSO DOES NOT WIPE THE SCOPE DOWN AFTER THE FLUSHING WITH DETERGENT SOLUTION AND DOES NOT LET IT SOAK FOR THE DETERGENT MANUFACTURER'S RECOMMENDED TIME. THE SCOPE DOES NOT SOAK AT ALL AND GOES DIRECTLY INTO THE RINSE SINK. THE CUSTOMER DOES NOT ALWAYS KEEP THE SCOPE IMMERSED WHILE REPROCESSING AND BRUSHING. THE CUSTOMER USES PURE DETERGENT. THE CUSTOMER USES THE SCOPE BUDDY PLUS FOR FLUSHING AND THE MEDIVATOR¿S EDGE AUTOMATIC ENDOSCOPY REPROCESSOR (AER). THE CORRECT REPROCESSING METHOD PER THE INSTRUCTIONS FOR USE (IFU) WERE REVIEWED IN THE IN-SERVICE. CUSTOMER MANAGER WAS PROVIDED WITH A COPY OF THE DOCUMENTATION OF THE IN-SERVICE, POSTERS, AND WEBSITES FOR REPROCESSING VIDEOS. CUSTOMER WAS ADVISED TO FOLLOW ALL VALIDATED REPROCESSING METHODS LISTED IN THE REPROCESSING IFU.
THIS REPORT IS BEING SUPPLEMENTED TO PROVIDE ADDITIONAL INFORMATION BASED ON THE LEGAL MANUFACTURER'S FINAL INVESTIGATION. A REVIEW OF THE DEVICE HISTORY RECORD FOUND NO DEVIATIONS THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED ISSUE. BASED ON THE RESULTS OF THE INVESTIGATION, THE ROOT CAUSE OF THE EVENT WAS UNABLE TO BE IDENTIFIED. IT WAS CONFIRMED THAT NO LINT-FREE CLOTH IS USED DURING MANUAL CLEANING, THERE WERE CONFIRMED DEVIATIONS FROM THE INSTRUCTIONS FOR USE (IFU). IT IS LIKELY THAT THERE WAS A DIFFERENCE IN THE HANDLING OF EQUIPMENT AND RECOGNITION OF REPROCESSING PROCEDURES BETWEEN OLYMPUS' RECOMMENDATIONS AND THE FACILITIES CONCERNED. IT WAS ALSO CONFIRMED THERE IS A PLACE IN THE SCOPE CHANNEL THAT THE CLEANING BRUSH DOES NOT HIT. THE EVENT CAN BE DETECTED/PREVENTED BY FOLLOWING THE INSTRUCTIONS FOR USE WHICH STATE: 1) "ALL CLOTHS USED IN REPROCESSING ARE RECOMMENDED TO BE LINT-FREE. LINT OR CLOTH FIBERS SHED INTO REPROCESSING FLUIDS MAY BE INJECTED INTO THE ENDOSCOPE CHANNELS. THERE IS THE POTENTIAL FOR LINT OR CLOTH FIBERS TO LODGE IN CHANNELS OR BECOME TRAPPED IN THE AIR/WATER NOZZLE. IF GAUZE IS USED TO REPROCESS THE ENDOSCOPE, ENSURE THAT FIBERS DO NOT GET CAUGHT ON OR REMAIN TRAPPED BY PROTRUDING COMPONENTS LIKE THE AIR/WATER NOZZLE." 2) 5.5 MANUALLY CLEANING THE ENDOSCOPE AND ACCESSORIES BRUSH THE CHANNELS DESCRIBES THE BRUSHING PROCEDURE. 3) 5.5 MANUALLY CLEANING THE ENDOSCOPE AND ACCESSORIES FILL A CLEAN, LARGE BASIN WITH THE DETERGENT SOLUTION AT THE TEMPERATURE AND CONCENTRATION RECOMMENDED BY THE DETERGENT MANUFACTURER. OLYMPUS WILL CONTINUE TO MONITOR FIELD PERFORMANCE FOR THIS DEVICE.
AS REPORTED FOR THIS EVENT BY THE CUSTOMER, CONCERN WAS RAISED FOR INSUFFICIENT REPROCESSING OF DEVICES ACROSS MULTIPLE CUSTOMER SITES UPON FINDINGS AFTER DEVICES WERE INSPECTED WITH A BORESCOPE. THE ACTUAL FINDINGS OF A BORESCOPE INSPECTION FOR THIS DEVICE IS LOOSE FIBERS IN DISTAL TIP. THERE IS NO REPORTED HARM TO ANY PATIENT OR PERSONS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 235231 | EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE | GASTROINTESTINAL VIDEOSCOPE | FDS | AIZU OLYMPUS CO., LTD. | GIF-H190 | 04953170305290 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |