FDA Adverse Event Malfunction Summary report: N

TITAN SGS

MDR report key: 15370883 · Received September 7, 2022

Report

Report Number
3012481535-2022-00006
Event Type
Malfunction
Date Received
September 7, 2022
Date of Event
August 8, 2022
Report Date
September 7, 2022
Manufacturer
STANDARD BARIATRICS
Product Code
GDW
UDI-DI
00851677007089
PMA / PMN Number
K210278
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

UPON EVALUATION OF THE TITAN SGS, IT WAS DETERMINED THAT THE LOAD APPLIED TO THE CLOSURE SCREW WAS EXCESSIVE AND CAUSED IT TO BEND/DISTORT DURING THE OPENING OF THE JAWS OF THE TITAN STAPLER, AS DAMAGE WAS OBSERVED. THE INVESTIGATION FOUND THAT THE DAMAGE TO THE CLOSURE SYSTEM LED TO THE DEVICE BEING INOPERABLE. THE ROOT CAUSE WAS ATTRIBUTED TO EXCESSIVE FORCE/ROTATIONS BEING APPLIED BY THE USER TO THE CLOSURE SYSTEM WHILE ATTEMPTING TO MANUALLY OPEN THE JAWS OF THE DEVICE USING THE BAILOUT KEY, RESULTING IN THE DEVICE BEING STUCK IN THE OPEN POSITION.

Description of Event or Problem · 0

DURING A SLEEVE GASTRECTOMY PROCEDURE, AT THE BEGINNING OF FIRING MODE, A HIGH LOAD WAS DETECTED WHICH ULTIMATELY LED TO THE USER REQUIRING TO ENGAGE IN MANUAL BAILOUT OF THE TITAN SGS STAPLER. DURING MANUAL BAIL OUT AND USING THE BAILOUT KEY PROVIDED, THE USER WAS ABLE TO OPEN THE TITAN SGS JAWS TO RELEASE FROM THE STOMACH. THE USER WAS UNABLE TO CLOSE THE DEVICE USING THE BAILOUT KEY, AS THE BAILOUT KEY SPUN FREELY, WHICH RESULTED IN REQUIRING THE TROCAR INCISION SITE TO BE SLIGHTLY INCREASED TO ALLOW FOR THE REMOVAL OF THE TITAN SGS WITH THE JAWS IN THE OPEN POSITION. THE PROCEDURE WAS PROLONGED, APPROXIMATELY 5 MINUTES, TO REMOVE THE DEVICE. THE PROCEDURE WAS COMPLETED USING A DIFFERENT STAPLER. THERE WAS NO ADDITIONAL PATIENT CONSEQUENCE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1717709 TITAN SGS SURGICAL STAPLER WITH STAPLE, IMPLANATABLE, PRODUCT CODE: GDW GDW STANDARD BARIATRICS SGS23R 103-22 00851677007089

Patients

Seq Age Sex Outcome Treatment
1 Unknown Required Intervention