FDA Adverse Event Death Summary report: N

INFUSOMAT SPACE

MDR report key: 15340971 · Received September 1, 2022

Report

Report Number
9610825-2022-00362
Event Type
Death
Date Received
September 1, 2022
Date of Event
August 17, 2022
Report Date
September 1, 2022
Manufacturer
B.BRAUN MELSUNGEN AG
Product Code
FRN
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BR
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THIS REPORT HAS BEEN IDENTIFIED AS B. BRAUN MELSUNGEN (B)(4) INTERNAL REPORT (B)(4). THE COMPLAINT IS UNDER EVALUATION. A FOLLOW-UP REPORT WILL BE PROVIDED AFTER THE EXAMINATION RESULTS ARE AVAILABLE.

Description of Event or Problem · 0

AS REPORTED BY THE USER FACILITY INFORMATION BY BBM SALES ORGANIZATION IN BRAZIL: "UNDERINFUSION (STANDSTILL WITHOUT ALARM)". "AS INFORMATION RECEIVED ON 2022/08/19 FROM CUSTOMER AND AFTER THE COMPLAINT REGISTRATION: "AT 16:19PM THE PUMP SHUTS OFF WITHOUT ALARM AND WITHOUT ANY KIND OF SIGNAL THAT IT WOULD STOP. SEVERAL EMPLOYEES PASS BY THE PATIENT'S BED, BUT NONE OF THEM NOTICE THAT IT IS TURNED OFF. ONLY ALMOST 1 HOUR LATER, THE NURSE RESPONSIBLE FOR THE PATIENT, AFTER BEING SIGNALED OF THE EVENT, NOTICES AND STARTS THE PUMP IMMEDIATELY. THE SAME THING WORKS AGAIN, AND THE NURSE MOVES BEHIND IT, NOT BEING ABLE TO SEE IF THERE WAS A CONNECTION BETWEEN THE PUMP AND THE POWER CABLE. THE PUMP REMAINS IN OPERATION FOR A FEW MOMENTS, UNTIL THE VERY UNSTABLE PATIENT PRESENTS CARDIORESPIRATORY ARREST AND RHYTHM OF ASYSTOLE, UNDERGOING A MANEUVER FOR 8', BEING CARDIOVERTED. THE PATIENT RETURNS IN A SERIOUS GENERAL CONDITION. IN ACCORDANCE WITH CUSTOMER: "PATIENT IN REFRACTORY SHOCK EVOLVING WITH CARDIORESPIRATORY ARREST AND PULSELESS ELECTRICAL ACTIVITY, USING VASOACTIVE AMINES AT MAXIMUM DOSE, UNRESPONSIVE TO MEASURES INSTITUTED TO CHANGE THE PROGNOSIS. CARDIOPULMONARY RESUSCITATION MANEUVERS NOT PERFORMED" PATIENT OUTCOME: DEATH ON (B)(6) 2022. "

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1809279 INFUSOMAT SPACE PUMP, INFUSION FRN B.BRAUN MELSUNGEN AG

Patients

Seq Age Sex Outcome Treatment
1 Unknown