INFUSOMAT SPACE
Report
- Report Number
- 9610825-2022-00362
- Event Type
- Death
- Date Received
- September 1, 2022
- Date of Event
- August 17, 2022
- Report Date
- September 1, 2022
- Manufacturer
- B.BRAUN MELSUNGEN AG
- Product Code
- FRN
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- BR
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
THIS REPORT HAS BEEN IDENTIFIED AS B. BRAUN MELSUNGEN (B)(4) INTERNAL REPORT (B)(4). THE COMPLAINT IS UNDER EVALUATION. A FOLLOW-UP REPORT WILL BE PROVIDED AFTER THE EXAMINATION RESULTS ARE AVAILABLE.
AS REPORTED BY THE USER FACILITY INFORMATION BY BBM SALES ORGANIZATION IN BRAZIL: "UNDERINFUSION (STANDSTILL WITHOUT ALARM)". "AS INFORMATION RECEIVED ON 2022/08/19 FROM CUSTOMER AND AFTER THE COMPLAINT REGISTRATION: "AT 16:19PM THE PUMP SHUTS OFF WITHOUT ALARM AND WITHOUT ANY KIND OF SIGNAL THAT IT WOULD STOP. SEVERAL EMPLOYEES PASS BY THE PATIENT'S BED, BUT NONE OF THEM NOTICE THAT IT IS TURNED OFF. ONLY ALMOST 1 HOUR LATER, THE NURSE RESPONSIBLE FOR THE PATIENT, AFTER BEING SIGNALED OF THE EVENT, NOTICES AND STARTS THE PUMP IMMEDIATELY. THE SAME THING WORKS AGAIN, AND THE NURSE MOVES BEHIND IT, NOT BEING ABLE TO SEE IF THERE WAS A CONNECTION BETWEEN THE PUMP AND THE POWER CABLE. THE PUMP REMAINS IN OPERATION FOR A FEW MOMENTS, UNTIL THE VERY UNSTABLE PATIENT PRESENTS CARDIORESPIRATORY ARREST AND RHYTHM OF ASYSTOLE, UNDERGOING A MANEUVER FOR 8', BEING CARDIOVERTED. THE PATIENT RETURNS IN A SERIOUS GENERAL CONDITION. IN ACCORDANCE WITH CUSTOMER: "PATIENT IN REFRACTORY SHOCK EVOLVING WITH CARDIORESPIRATORY ARREST AND PULSELESS ELECTRICAL ACTIVITY, USING VASOACTIVE AMINES AT MAXIMUM DOSE, UNRESPONSIVE TO MEASURES INSTITUTED TO CHANGE THE PROGNOSIS. CARDIOPULMONARY RESUSCITATION MANEUVERS NOT PERFORMED" PATIENT OUTCOME: DEATH ON (B)(6) 2022. "
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1809279 | INFUSOMAT SPACE | PUMP, INFUSION | FRN | B.BRAUN MELSUNGEN AG |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |