FDA Adverse Event Malfunction Summary report: N

PENTAX

MDR report key: 15332885 · Received September 1, 2022

Report

Report Number
9610877-2022-57310
Event Type
Malfunction
Date Received
September 1, 2022
Date of Event
August 9, 2022
Report Date
September 1, 2022
Manufacturer
HOYA CORPORATION PENTAX TOKYO OFFICE
Product Code
PEA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
HR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

THIS DEVICE IS CLASSIFIED AS IMPORT FOR EXPORT, THEREFORE 510K IS NOT APPLICABLE. MODEL EPK-3000-US IS AVAILABLE IN THE USA WITH A 510K NUMBER K172156. WE CHECKED THE RETURNED UNIT AND CONFIRMED THAT THE IRIS DIAPHRAGM MALFUNCTION. BASED ON THE RESULT, WE CONCLUDED THAT IT WAS CAUSED DUE TO THE EXCESSIVE SHOCK APPLIED ON THE IRIS DIAPHRAGM. IN ADDITION, WE CONFIRMED THAT THE BATTERY WEAK, THE PRE-PROCESS PCB MALFUNCTION, AND THE XENON LAMP WORN OUT; HOWEVER, THEY ARE NOT THE MAIN CAUSE, AND/OR IRRELEVANT TO THE ALLEGED COMPLAINT. BASED ON THE TECHNICAL REPORT AND/OR THE RISK ANALYSIS RESULTS, IT WAS EVALUATED TO SUBMIT MDR.

Description of Event or Problem · 0

THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. IRIS DIAPHRAGM FAILURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2353606 PENTAX VIDEO PROCESSOR PEA HOYA CORPORATION PENTAX TOKYO OFFICE EPK-I

Patients

Seq Age Sex Outcome Treatment
1 Unknown