FDA Adverse Event Malfunction Summary report: N

3090 SP SURGICAL TABLE

MDR report key: 15329150 · Received August 31, 2022

Report

Report Number
1043572-2022-00058
Event Type
Malfunction
Date Received
August 31, 2022
Date of Event
August 3, 2022
Report Date
August 31, 2022
Manufacturer
STERIS CORPORATION - MONTGOMERY
Product Code
FWW
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

A STERIS SERVICE TECHNICIAN INSPECTED THE TABLE AND FOUND HYDRAULIC OIL IN THE BOTTOM OF THE TABLE BASE AND THE HYDRAULIC OIL LEVEL TO BE LOW. THE TECHNICIAN ADDED HYDRAULIC OIL, TESTED THE TABLE, AND FOUND IT TO BE OPERATING PROPERLY. A FEW DAYS FOLLOWING THE TECHNICIAN'S INITIAL INSPECTION, HE PERFORMED PREVENTIVE MAINTENANCE AND CONFIRMED THE TABLE TO BE OPERATING TO SPECIFICATIONS. THE SURGICAL TABLE WAS RETURNED TO SERVICE. THE 3080SP SUBJECT OF THE REPORTED EVENT IS APPROXIMATELY 28 YEARS OLD AND IS SERVICED AND MAINTAINED BY THE USER FACILITY. NO ADDITIONAL ISSUES HAVE BEEN REPORTED.

Description of Event or Problem · 0

THE USER FACILITY REPORTED THAT DURING A PATIENT PROCEDURE AN EMPLOYEE COMMANDED THEIR 3080SP SURGICAL TABLE TO RAISE IN HEIGHT HOWEVER, THE TABLE "SHUTTERED" IN A DOWNWARD MOVEMENT AND MADE A "LOUD NOISE". THE PROCEDURE WAS COMPLETED SUCCESSFULLY, AND NO INJURIES WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2305344 3090 SP SURGICAL TABLE SURGICAL TABLE FWW STERIS CORPORATION - MONTGOMERY

Patients

Seq Age Sex Outcome Treatment
1 Unknown