FDA Adverse Event Malfunction Summary report: N

INRATIO

MDR report key: 1532805 · Received October 29, 2009

Report

Report Number
2027969-2009-00982
Event Type
Malfunction
Date Received
October 29, 2009
Date of Event
October 5, 2008
Report Date
October 29, 2009
Manufacturer
BIOSITE INCORPORATED
Product Code
JPA
PMA / PMN Number
021923
Removal / Correction Number
NI
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION: STRIP CODE: 4X2YN, STRIP LOT: 080818. PER TECH SUPPORT REP, THE FIRST TWO TESTS WERE CONDUCTED OVER AN HOUR APART. STRIP CODE: MY55J, STRIP LOT: 213037. THE MEAN OF THE INRATIO METER AND COMPARATIVE SYSTEM INR WERE CALCULATED. BOTH INRATIO AND REFERENCE VALUES ARE WITHIN THE CONFIDENCE LIMITS FOR INR TESTING. THE RESULTS ARE NOT CONSIDERED DISCREPANT WITHIN THE CONTEXT OF THE DOCUMENTED VARIABILITY FOR INR TESTING. NO FURTHER INVESTIGATION IS REQUIRED. SUMMARY: MEAN CALCULATION FROM COMPARISON TEST DATA PROVIDED BY END-USER REVEALED BOTH INRATIO AND LAB VALUES ARE WITHIN THE CONFIDENCE LIMITS FOR INR TESTING. THE RESULTS ARE NOT CONSIDERED DISCREPANT WITHIN THE CONTEXT OF THE DOCUMENTED VARIABILITY FOR INR TESTING. METER WAS EXPECTED TO BE RETURNED BUT NOT RECEIVED. FURTHER TEST IS NOT REQUIRED AT THIS TIME. AS OF 10/13/2009, 4 DISCREPANT RESULTS COMPLAINTS WERE REPORTED FOR LOT # 213037 YIELDING A COMPLAINT RATE OF 0.008%. DUE TO THIS LOW OCCURRENCE RATE, BELOW THE ACTION THRESHOLDS MONITORED BY QA FOR CORRECTIVE ACTION (>0.07%) NO FURTHER ACTION IS REQUIRED AT THIS TIME. ON GOING TRENDING SHALL BE PERFORMED TO DETERMINE IF FURTHER ACTION IS WARRANTED. AS OF 10/29/2009, 22 DISCREPANT RESULTS COMPLAINTS WERE REPORTED FOR LOT# 080818 YIELDING A COMPLAINT RATE OF 0.006%. DUE TO THIS LOW OCCURRENCE RATE, BELOW THE ACTION THRESHOLDS MONITORED BY QA FOR CORRECTIVE ACTION (>0.07%) NO FURTHER ACTION IS REQUIRED AT THIS TIME. ON GOING TRENDING SHALL BE PERFORMED TO DETERMINE IF FURTHER ACTION IS WARRANTED. NO CORRECTIVE ACTION IS REQUIRED AT THIS TIME AS NO PRODUCT DISCREPANCY WAS ESTABLISHED.

Description of Event or Problem · 1

"CALLER ALLEGED DISCREPANT RESULTS COMPARED WITH THE LAB. RESULTS AS FOLLOWS:" USING STRIP CODE: 4X2YN. USING STRIP CODE: MY55J.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INRATIO PROTHROMBIN TIME TEST JPA BIOSITE INCORPORATED 0100267 080818

Patients

Seq Age Sex Outcome Treatment
1 NI