EREADER+ HOST PC
Report
- Report Number
- 3003110749-2022-00007
- Event Type
- Malfunction
- Date Received
- August 30, 2022
- Date of Event
- August 2, 2022
- Report Date
- August 2, 2022
- Manufacturer
- AMEDITECH, INC.
- Product Code
- LNX
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
THE DEVICE IN QUESTION IS A PERIPHERAL HARDWARE UNIT USED IN CONJUNCTION WITH OUR EREADER+ DEVICE. THERE WAS NO DEFICIENCY NOTED WITH THE EREADER+ THAT WAS CONNECTED TO THE COMPUTER. INVESTIGATION OF THIS EVENT IS IN PROGRESS. A FOLLOW-UP REPORT WILL BE SUBMITTED UPON COMPLETION OF THE INVESTIGATION.
THE COMPLAINT PRODUCT WAS NOT RETURNED. THE THIRD-PARTY MANUFACTURER OF THE COMPUTER HAS BEEN MADE AWARE OF THIS POWER SUPPLY ISSUE FOR INVESTIGATION. A MEDITECH WILL CONTINUE TO MONITOR AND TREND THIS ISSUE.
IT WAS REPORTED THAT THEIR EREADER COMPUTER HAD A BLACK SCREEN. DISCONNECTING AND RECONNECTING THE SYSTEM, AS ADVISED BY THE TECHNICAL SERVICES GROUP, WAS UNSUCCESSFUL IN RESOLVING THE PROBLEM. DURING THIS TROUBLESHOOTING SESSION, IT WAS DISCOVERED THAT THE POWER SUPPLY WIRES WERE EXTREMELY HOT. THE CUSTOMER THEN NOTED THAT THEIR UNIT HAD COMPLETELY DIED. UPON FOLLOWING UP FOR ADDITIONAL INFORMATION, IT WAS CONFIRMED THAT THERE WERE NO INJURIES THAT RESULTED FROM THE OVERHEATING WIRES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 332337 | EREADER+ HOST PC | COMPUTER | LNX | AMEDITECH, INC. | M720Q |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |