FDA Adverse Event Malfunction Summary report: N

EREADER+ HOST PC

MDR report key: 15318517 · Received August 30, 2022

Report

Report Number
3003110749-2022-00007
Event Type
Malfunction
Date Received
August 30, 2022
Date of Event
August 2, 2022
Report Date
August 2, 2022
Manufacturer
AMEDITECH, INC.
Product Code
LNX
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE DEVICE IN QUESTION IS A PERIPHERAL HARDWARE UNIT USED IN CONJUNCTION WITH OUR EREADER+ DEVICE. THERE WAS NO DEFICIENCY NOTED WITH THE EREADER+ THAT WAS CONNECTED TO THE COMPUTER. INVESTIGATION OF THIS EVENT IS IN PROGRESS. A FOLLOW-UP REPORT WILL BE SUBMITTED UPON COMPLETION OF THE INVESTIGATION.

Additional Manufacturer Narrative · 0

THE COMPLAINT PRODUCT WAS NOT RETURNED. THE THIRD-PARTY MANUFACTURER OF THE COMPUTER HAS BEEN MADE AWARE OF THIS POWER SUPPLY ISSUE FOR INVESTIGATION. A MEDITECH WILL CONTINUE TO MONITOR AND TREND THIS ISSUE.

Description of Event or Problem · 0

IT WAS REPORTED THAT THEIR EREADER COMPUTER HAD A BLACK SCREEN. DISCONNECTING AND RECONNECTING THE SYSTEM, AS ADVISED BY THE TECHNICAL SERVICES GROUP, WAS UNSUCCESSFUL IN RESOLVING THE PROBLEM. DURING THIS TROUBLESHOOTING SESSION, IT WAS DISCOVERED THAT THE POWER SUPPLY WIRES WERE EXTREMELY HOT. THE CUSTOMER THEN NOTED THAT THEIR UNIT HAD COMPLETELY DIED. UPON FOLLOWING UP FOR ADDITIONAL INFORMATION, IT WAS CONFIRMED THAT THERE WERE NO INJURIES THAT RESULTED FROM THE OVERHEATING WIRES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
332337 EREADER+ HOST PC COMPUTER LNX AMEDITECH, INC. M720Q

Patients

Seq Age Sex Outcome Treatment
1 Unknown