IMRIS ORT300 OPERATING ROOM TABLE
Report
- Report Number
- 3010326005-2022-00014
- Event Type
- Malfunction
- Date Received
- August 29, 2022
- Date of Event
- July 30, 2022
- Report Date
- July 30, 2022
- Manufacturer
- IMRIS - DEERFIELD IMAGING, INC.
- Product Code
- KXJ
- UDI-DI
- 00857534006356
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- 003
Narratives
THE SUSPECTED SOURCE OF THE FLUID LEAK IS A VALVE COMPONENT WITHIN THE HYDRAULIC MANIFOLD. NO HYDRAULIC PRESSURE LOSS OR IMPACT ON TABLE FUNCTION HAS BEEN OBSERVED. A FOLLOWUP REPORT WILL BE SUBMITTED ONCE THE INVESTIGATION IS COMPLETE.
AN IMRIS CUSTOMER SERVICE ENGINEER MADE MULTIPLE SITE VISITS TO ADDRESS THE PREVENTIVE MAINTENANCE FINDING OF A SMALL HYDRAULIC FLUID LEAK FROM THE OPERATING ROOM TABLE. THE SOURCE OF THE LEAK WAS NOT DEFINITIVELY DETERMINED, AT A FOLLOWUP VISIT NO FLUID OUTSIDE THE HYDRAULIC SYSTEM WAS OBSERVED. MULTIPLE SUSPECTED HYDRAULIC SYSTEM COMPONENTS WERE REPLACED AS A PREVENTIVE MEASURE, INCLUDING A FLOW CONTROL VALVE, A HOSE, AND MULTIPLE FITTINGS AND UNIONS. HYDRAULIC SYSTEM PRESSURE AND FLUID LEVELS WERE MEASURED TO BE WITHIN SPECIFICATION, AND WEIGHTED DRIFT TESTING WAS PERFORMED WITH PASSING RESULTS. NO IMPACT TO TABLE FUNCTION WAS REPORTED OR OBSERVED.
DURING ROUTINE PREVENTIVE MAINTENANCE, AN IMRIS CUSTOMER SERVICE ENGINEER OBSERVED SEVERAL SMALL DROPS OF HYDRAULIC FLUID ON OUTSIDE OF THE HYDRAULIC LINES OF THE ORT300 OPERATING ROOM TABLE. NO ISSUES WERE REPORTED OR OBSERVED WITH RESPECT TO HYDRAULICS CONTROLLED TABLE FUNCTIONS. THERE WAS NO PATIENT INVOLVEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 243173 | IMRIS ORT300 OPERATING ROOM TABLE | OPERATING ROOM TABLE | KXJ | IMRIS - DEERFIELD IMAGING, INC. | ORT300 | 00857534006356 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |