COBAS 8000 C 702 MODULE
Report
- Report Number
- 1823260-2022-02617
- Event Type
- Malfunction
- Date Received
- August 29, 2022
- Date of Event
- July 21, 2022
- Report Date
- September 14, 2022
- Manufacturer
- ROCHE DIAGNOSTICS
- Product Code
- JJE
- UDI-DI
- 04015630930845
- PMA / PMN Number
- K100853
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IT
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
IT IS UNKNOWN WHETHER THE MEDICATION OF VITAMIN B 6 WAS GIVEN AFTER THE SAMPLE WAS TAKEN OR BEFORE. IT IS ALSO NOT KNOWN WHETHER THE MEDICATION IS CONTINUOUSLY TAKEN. A RE-RUN OF THE SAME SAMPLE ON COBAS 8000 C 702 MODULE RESULTED IN A VALUE OF 37.9 MOL/L. CALIBRATION DATA WERE WITHIN THE EXPECTED RANGES. QUALITY CONTROL RECOVERY WAS WITHIN SPECIFICATION ON THE DATE OF THE EVENT. UPON REVIEW OF THE ALARM TRACE FROM 21-JUL-2022, THERE WERE MULTIPLE ABNORMAL ASPIRATION ALARMS. THE INVESTIGATION COULD NOT IDENTIFY A PRODUCT PROBLEM. THE ISSUE IS CONSISTENT WITH INCORRECT PRE-ANALYTIC HANDLING OF THE SAMPLE.
THE INITIAL REPORTER STATED THEY RECEIVED DISCREPANT RESULTS FOR ONE PATIENT SAMPLE TESTED WITH THE HCYS (HOMOCYSTEINE ENZYMATIC) ASSAY ON A COBAS 8000 C 702 MODULE COMPARED TO THE RESULT OBTAINED WITH HPLC (HIGH-PERFORMANCE LIQUID CHROMATOGRAPHY). THE SAMPLE INITIALLY RESULTED IN AN HYCS VALUE OF 7.6 MOL/L. REPETITION OF THE SPECIMEN CONFIRMED THIS VALUE. THE SAMPLE WAS MEASURED USING THE HPLC TECHNIQUE, RESULTING IN A VALUE OF 40 MOL/L. NO ERRONEOUS RESULT WAS REPORTED OUTSIDE OF THE LABORATORY. THE HYCS REAGENT LOT NUMBER WAS 571023, WITH AN EXPIRATION DATE OF 31-AUG-2022.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2344116 | COBAS 8000 C 702 MODULE | CLINICAL CHEMISTRY ANALYZER | JJE | ROCHE DIAGNOSTICS | C702 | NA | 04015630930845 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |