FDA Adverse Event Other Summary report: N

COMLINX NCM SYSTEM

MDR report key: 1530960 · Received October 22, 2009

Report

Report Number
2027454-2009-00020
Event Type
Other
Date Received
October 22, 2009
Date of Event
November 18, 2008
Report Date
October 15, 2009
Manufacturer
HILL-ROM MANUFACTURING, INC., HILL-ROM IT SOLUTIONS
Product Code
ILQ
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(PURSUANT TO THE RESPONSE TO WARNING LETTER 2008-DT-04 ISSUED TO HILL-ROM ((B)(4)), HILL-ROM HAS EXTENDED ITS RETROSPECTIVE REVIEW OF EVENTS FOR REPORTABILITY TO INCLUDE THE (B)(4) FACILITY. THIS EFFORT HAS IDENTIFIED THE FOLLOWING EVENT AS A REPORTABLE EVENT. THUS, A MDR IS BEING SUBMITTED TO FULFILL REPORTING REQUIREMENTS.) INVESTIGATION OF THE SYSTEM CALL REPORTS AND STATUS REPORTS INDICATE THAT APPROX 30 MINUTES PRIOR TO THE ALLEGED PT FALL, A BED EXIT CALL WAS PLACED, THEN STAFF WAS LOCATED IN THE ROOM, AND THEN BED EXIT NOTIFICATION WAS TURNED OFF. THE BED EXIT NOTIFICATION WAS NOT TURNED BACK ON UNTIL APPROX 90 MINUTES AFTER THE PT FALL. TECHNICIAN EVALUATED AND TESTED THE SYSTEM, NO SYSTEM PROBLEMS OR MALFUNCTIONS WERE FOUND. THE TECHNICIAN WAS NOT ABLE TO REPRODUCE ANY ISSUE WHERE THE ALERTS DID NOT WORK PROPERLY OR WHERE THE BED EXIT DISABLED ITSELF AUTOMATICALLY. INVESTIGATION RESULTS SUGGESTS THIS IS A RESULT OF A USER INTERFACE ERROR IN HOW TO PROPERLY ENABLE AND DISABLE THE BED EXIT WHEN NURSING STAFF ENTER THE ROOM AND ARE TAKING A PT OUT OF BED. ADD'L CLINICAL TRAINING ON THE USE OF CARE ALERTS WAS PROVIDED.

Description of Event or Problem · 1

CUSTOMER INDICATED A PT EXITED THE BED AND FELL, AND THERE WAS NO BED EXIT ALERT. WHEN THIS OCCURRED THE NURSE INDICATED SHE HAD PREVIOUSLY CONFIRMED THE BED EXIT WAS ON, BUT WHEN CHECKED AFTER THE EVENT, THE BED EXIT WAS OFF, SHE STATES THIS WAS WITH NO HUMAN INTERVENTION. THE PT DID RECEIVE MINOR BRUISES AND A CUT ON FACE, BUT PT OUTCOME WAS OK.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COMLINX NCM SYSTEM COMMUNICATIONS ILQ HILL-ROM MANUFACTURING, INC., HILL-ROM IT SOLUTIONS P250001 NCM V4.0

Patients

Seq Age Sex Outcome Treatment
1 NA