COMLINX NCM SYSTEM
Report
- Report Number
- 2027454-2009-00020
- Event Type
- Other
- Date Received
- October 22, 2009
- Date of Event
- November 18, 2008
- Report Date
- October 15, 2009
- Manufacturer
- HILL-ROM MANUFACTURING, INC., HILL-ROM IT SOLUTIONS
- Product Code
- ILQ
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER
Narratives
(PURSUANT TO THE RESPONSE TO WARNING LETTER 2008-DT-04 ISSUED TO HILL-ROM ((B)(4)), HILL-ROM HAS EXTENDED ITS RETROSPECTIVE REVIEW OF EVENTS FOR REPORTABILITY TO INCLUDE THE (B)(4) FACILITY. THIS EFFORT HAS IDENTIFIED THE FOLLOWING EVENT AS A REPORTABLE EVENT. THUS, A MDR IS BEING SUBMITTED TO FULFILL REPORTING REQUIREMENTS.) INVESTIGATION OF THE SYSTEM CALL REPORTS AND STATUS REPORTS INDICATE THAT APPROX 30 MINUTES PRIOR TO THE ALLEGED PT FALL, A BED EXIT CALL WAS PLACED, THEN STAFF WAS LOCATED IN THE ROOM, AND THEN BED EXIT NOTIFICATION WAS TURNED OFF. THE BED EXIT NOTIFICATION WAS NOT TURNED BACK ON UNTIL APPROX 90 MINUTES AFTER THE PT FALL. TECHNICIAN EVALUATED AND TESTED THE SYSTEM, NO SYSTEM PROBLEMS OR MALFUNCTIONS WERE FOUND. THE TECHNICIAN WAS NOT ABLE TO REPRODUCE ANY ISSUE WHERE THE ALERTS DID NOT WORK PROPERLY OR WHERE THE BED EXIT DISABLED ITSELF AUTOMATICALLY. INVESTIGATION RESULTS SUGGESTS THIS IS A RESULT OF A USER INTERFACE ERROR IN HOW TO PROPERLY ENABLE AND DISABLE THE BED EXIT WHEN NURSING STAFF ENTER THE ROOM AND ARE TAKING A PT OUT OF BED. ADD'L CLINICAL TRAINING ON THE USE OF CARE ALERTS WAS PROVIDED.
CUSTOMER INDICATED A PT EXITED THE BED AND FELL, AND THERE WAS NO BED EXIT ALERT. WHEN THIS OCCURRED THE NURSE INDICATED SHE HAD PREVIOUSLY CONFIRMED THE BED EXIT WAS ON, BUT WHEN CHECKED AFTER THE EVENT, THE BED EXIT WAS OFF, SHE STATES THIS WAS WITH NO HUMAN INTERVENTION. THE PT DID RECEIVE MINOR BRUISES AND A CUT ON FACE, BUT PT OUTCOME WAS OK.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COMLINX NCM SYSTEM | COMMUNICATIONS | ILQ | HILL-ROM MANUFACTURING, INC., HILL-ROM IT SOLUTIONS | P250001 | NCM V4.0 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |