PRIMEWIRE
Report
- Report Number
- 2939520-2009-00032
- Event Type
- Other
- Date Received
- November 16, 2009
- Date of Event
- October 22, 2009
- Report Date
- October 26, 2009
- Manufacturer
- VOLCANO CORP
- Product Code
- DXO
- PMA / PMN Number
- K070487
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AZ, US
- Reporter Occupation
- OTHER
Narratives
UPON RECEIPT, VISUAL INSPECTION SHOWED THAT THE WIRE WAS RECEIVED IN TWO PIECES. IT WAS RETURNED IN A PLASTIC BIOHAZARD BAG WITHOUT THE PROTECTIVE COIL. AS INDICATED IN THE COMPLAINT DESCRIPTION, THE WIRE WAS CUT INTO TWO PIECES. THE WIRE WAS CUT AT 37.5 CM FROM THE PROXIMAL END (CONNECTOR) OF THE WIRE. MULTIPLE KINKS WERE OBSERVED ON THE PROXIMAL AND DISTAL END OF THE WIRE. THE WIRE WAS MICROSCOPICALLY EXAMINED AND ALL COMPONENTS WERE INTACT AND PRESENT. THE PRESSURE SENSOR WAS INSPECTED AND YELLOW BROWN RUSTY RESIDUE WAS OBSERVED IN THE SENSOR HOUSING. SINCE THE DEVICE WAS CUT INTO TWO PIECES, THE FUNCTIONAL TESTING OF THE DEVICE COULD NOT BE PERFORMED. THEREFORE, WE WERE UNABLE TO DUPLICATE THE ORIGINAL COMPLAINT OF 107 ERROR MESSAGE. BASED ON THE COMPLAINT DESCRIPTION, THE WIRE BECAME BENT WHEN PHYSICIAN WAS TRYING TO LOAD THE BALLOON CATHETER AND THEN WIRE WAS CUT BY THE PHYSICIAN. ALTHOUGH THERE WAS NO PT IMPACT DUE TO THIS INCIDENT, THERE IS A POTENTIAL FOR INJURY SHOULD THIS HAPPEN AGAIN. THIS REPORT IS BEING SENT AS A NOTIFICATION. (B)(4).
FFR OF PLB 0.39 RESTING. PULLED WIRE BACK TO ASSESS MID RCA LESION. GOT NO SIGNAL (WIRE PRESS) (ERROR 107) EVEN THOUGH WIRE WAS REATTACHED SEVERAL TIMES. HAD TO PULL AN IVUS CATHETER THEN STENTED MID RCA. WHEN ATTEMPTING TO LOAD NC BALLOON, WIRE BENT AND WAS CUT BY DR. THEN NC BALLOON LOADED ON CUT PRIMEWIRE. IT RUPTURED THE BALLOON. WE THEN HAD TO BUDDY A BMW, PULL THE PRIMEWIRE AND USE A SECOND NC BALLOON.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PRIMEWIRE | TRANSDUCER, PRESSURE, CATHETER TIP | DXO | VOLCANO CORP | 7900 | 118 02573 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |