COOK CELECT® PLATINUM NAVALIGN JUGULAR VENA CAVA FILTER SET
Report
- Report Number
- 3002808486-2022-00928
- Event Type
- Malfunction
- Date Received
- August 26, 2022
- Report Date
- September 30, 2022
- Manufacturer
- WILLIAM COOK EUROPE
- Product Code
- DTK
- UDI-DI
- 10827002343099
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
OCCUPATION: NON-HEALTHCARE PROFESSIONAL. INVESTIGATION: IT HAS NOT BEEN POSSIBLE TO FURTHER INVESTIGATE OR EVALUATE THIS ALLEGED EVENT BASED ON THE LIMITED INFORMATION AND/OR NO DEVICE FAILURE PROVIDED TO DATE. CATALOG NUMBER AND LOT NUMBER ARE UNKNOWN, HOWEVER, THE ALLEGED CELECT- PT IS MANUFACTURED AND INSPECTED ACCORDING TO SPECIFICATIONS. NO EVIDENCE TO SUGGEST THAT THIS DEVICE WAS NOT MANUFACTURED ACCORDING TO SPECIFICATIONS AND NOTHING INDICATES THAT THE FILTER DID NOT PERFORM AS INTENDED, E.G. INTENDED FOR THE PREVENTION OF RECURRENT PULMONARY EMBOLISM (PE) VIA PLACEMENT IN THE VENA CAVA. COOK WILL REOPEN ITS INVESTIGATION IF FURTHER INFORMATION IS RECEIVED WARRANTING SUPPLEMENTATION IN ACCORDANCE WITH 21 C.F.R. 803.56. THIS REPORT INCLUDES INFORMATION KNOWN AT THIS TIME. A FOLLOW-UP MEDWATCH REPORT WILL BE SUBMITTED IF ADDITIONAL RELEVANT INFORMATION BECOMES AVAILABLE. THIS REPORT IS REQUIRED BY THE FDA UNDER 21 CFR PART 803. THIS REPORT IS BASED ON UNCONFIRMED INFORMATION SUBMITTED BY OTHERS. NEITHER THE SUBMISSION OF THIS REPORT NOR ANY STATEMENT MADE IN IT IS INTENDED TO BE AN ADMISSION THAT ANY COOK DEVICE IS DEFECTIVE OR MALFUNCTIONED, THAT A DEATH OR SERIOUS INJURY OCCURRED, OR THAT ANY COOK DEVICE CAUSED OR CONTRIBUTED TO OR IS LIKELY TO CAUSE OR CONTRIBUTE TO A DEATH OR SERIOUS INJURY IF A MALFUNCTION OCCURRED.
G4 (510K): K211875. INVESTIGATION: INVESTIGATION IS REOPENED DUE TO ADDITIONAL INFORMATION PROVIDED. THE FOLLOWING ALLEGATIONS HAVE BEEN INVESTIGATED: DEEP VEIN THROMBOSIS (DVT), DEVICE UNABLE TO BE RETRIEVE, PAIN, STRESS. THE REPORTED ALLEGATIONS HAVE BEEN FURTHER INVESTIGATED BASED ON THE INFORMATION PROVIDED TO DATE. IVC OCCLUSION/ THROMBOSIS, NEW DVT, IVC STENOSIS AS A REPORTED COMPLICATION, IS A KNOWN RISK IN RELATION TO FILTER IMPLANT AND IS WELL DOCUMENTED IN THE CLINICAL LITERATURE AND IN CLINICAL PRACTICE GUIDELINES. THIS IS SUPPORTED BY THE CLINICAL EVIDENCE REPORT ESTABLISHED TO ASSESS AVAILABLE CLINICAL DATA TO IDENTIFY AND EVALUATE THE CLINICAL SAFETY AND PERFORMANCE OF THE COOK VENA CAVA FILTERS. POTENTIAL ADVERSE EVENTS THAT MAY OCCUR INCLUDE, BUT ARE NOT LIMITED TO, THE FOLLOWING: VENA CAVA OCCLUSION OR THROMBOSIS, VENA CAVA STENOSIS, DEEP VEIN THROMBOSIS. PHYSICIAN PRACTICE GUIDELINES AND PUBLISHED GUIDANCE FROM REGULATORY AGENCIES RECOMMEND THAT PATIENTS WITH INDWELLING FILTERS UNDERGO ROUTINE FOLLOW-UP. THE RISKS/BENEFITS OF FILTER RETRIEVAL SHOULD BE CONSIDERED FOR EACH PATIENT DURING FOLLOW-UP. ONCE PROTECTION FROM PE IS NO LONGER NECESSARY, FILTER RETRIEVAL SHOULD BE CONSIDERED. FILTER RETRIEVAL SHOULD BE ATTEMPTED WHEN FEASIBLE AND CLINICALLY INDICATED. FILTER RETRIEVAL IS A PATIENT-SPECIFIC, CLINICALLY COMPLEX DECISION; THE DECISION TO REMOVE A FILTER SHOULD BE BASED ON EACH PATIENT¿S INDIVIDUAL RISK/BENEFIT PROFILE (E.G., A PATIENT¿S CONTINUED NEED FOR PROTECTION FROM PE COMPARED TO THEIR EXPERIENCE WITH AND (OR) ONGOING RISK OF EXPERIENCING FILTER-RELATED COMPLICATIONS). FOR ALL RETRIEVABLE IVC FILTERS, RETRIEVAL BECOMES MORE CHALLENGING WITH TIME, AND THIS IS COMMONLY DUE TO ENCAPSULATION OF THE FILTER LEGS OR HOOK (IN A TILTED FILTER) BY TISSUE INGROWTH. THE FILTER IS DESIGNED TO BE RETRIEVED WITH THE GÜNTHER TULIP VENA CAVA FILTER RETRIEVAL SET. IT MAY ALSO BE RETRIEVED WITH THE CLOVERSNARE® VASCULAR RETRIEVER. COOK HAS NOT PERFORMED TESTING TO EVALUATE THE SAFETY OR EFFECTIVENESS OF FILTER RETRIEVAL USING OTHER RETRIEVAL SYSTEMS OR TECHNIQUES. THE PUBLISHED CLINICAL LITERATURE INCLUDES DESCRIPTIONS OF ALTERNATIVE TECHNIQUES FOR FILTER RETRIEVAL; USE OF THESE TECHNIQUES VARIES ACCORDING TO PHYSICIAN EXPERIENCE, PATIENT ANATOMY, AND FILTER POSITION. THE SAFETY OR EFFECTIVENESS OF THESE ALTERNATIVE RETRIEVAL TECHNIQUES HAS NOT BEEN ESTABLISHED. SPECIFIC FOR ¿EMBEDDED¿ A FILTER THAT IS EMBEDDED IN THE WALL OF THE IVC MAY BE DIFFICULT TO RETRIEVE. FOR ALL RETRIEVABLE IVC FILTERS, RETRIEVAL BECOMES MORE CHALLENGING WITH TIME, AND THIS IS COMMONLY DUE TO ENCAPSULATION OF THE FILTER LEGS OR HOOK (IN A TILTED FILTER) BY TISSUE INGROWTH. UNKNOWN IF THE REPORTED PAIN AND STRESS ARE DIRECTLY RELATED TO THE FILTER AND UNABLE TO IDENTIFY A CORRESPONDING FAILURE MODE AT THIS POINT IN TIME. (B)(4) DEVICES IN LOT. NO RELEVANT NOTES ON WO FOR NEITHER DEVICE (IGTCFS-65-1-JUG-CELECT-PT) LOT: E3643753, NOR FILTER (IGTF-30-CELECT-PT) LOT: E3642271 AND E3641103. NO OTHER COMPLAINTS ON LOTS. PRODUCT IS MANUFACTURED AND INSPECTED ACCORDING TO SPECIFICATIONS. NO EVIDENCE TO SUGGEST THAT THIS DEVICE WAS NOT MANUFACTURED ACCORDING TO SPECIFICATIONS AND NOTHING INDICATES THAT THE FILTER DID NOT PERFORM AS INTENDED, E.G. INTENDED FOR THE PREVENTION OF RECURRENT PULMONARY EMBOLISM (PE) VIA PLACEMENT IN THE VENA CAVA. COOK WILL REOPEN ITS INVESTIGATION IF FURTHER INFORMATION IS RECEIVED WARRANTING SUPPLEMENTATION IN ACCORDANCE WITH 21 C.F.R. 803.56. THIS REPORT INCLUDES INFORMATION KNOWN AT THIS TIME. A FOLLOW-UP REPORT WILL BE SUBMITTED SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE. THIS REPORT IS REQUIRED BY THE FDA UNDER 21 CFR PART 803 AND IS BASED ON UNCONFIRMED INFORMATION SUBMITTED BY OTHERS. NEITHER THE SUBMISSION OF THIS REPORT NOR ANY STATEMENT CONTAINED HEREIN IS INTENDED TO BE AN ADMISSION THAT ANY COOK DEVICE IS DEFECTIVE OR MALFUNCTIONED, THAT A DEATH OR SERIOUS INJURY OCCURRED, NOR THAT ANY COOK DEVICE CAUSED, CONTRIBUTED TO, OR IS LIKELY TO CAUSE OR CONTRIBUTE TO A DEATH OR SERIOUS INJURY IF A MALFUNCTION OCCURRED.
IT IS ALLEGED THAT THE PATIENT RECEIVED A CELECT PLATINUM (PT) INFERIOR VENA CAVA (IVC) FILTER ON (B)(6) 2018, AND THE PATIENT WAS INJURED WITHOUT FURTHER EXPLANATION. HOSPITAL AND MEDICAL RECORDS HAVE BEEN REQUESTED, BUT NOT YET PROVIDED.
PATIENT ALLEGEDLY RECEIVED AN IMPLANT ON (B)(6) 2018 DUE TO DEEP VEIN THROMBOSIS (DVT)/ PULMONARY EMBOLISM (PE). PATIENT IS ALLEGING DEVICE IS UNABLE TO BE RETRIEVED. PATIENT FURTHER ALLEGES "AFTER HOURS OF THE SURGEON TRYING TO RETRIEVE THE DEVICE, I FELT IT OCCASIONALLY WITH PAIN. 6 WEEKS AFTER THE PROCEDURE I ENDED UP HAVING HEART SURGERY. THERE WAS A LOT OF STRESS WITH TRYING TO RETRIEVE THE DEVICE.". (B)(6) 2018: UNSUCCESSFUL RETRIEVAL REPORT PER RETRIEVAL REPORT (ATTEMPTED), "FLUOROSCOPIC GUIDED IVC VENOGRAM REVEALED A 3.6 X 1.6 CM THROMBUS PROJECTING THROUGH THE UPPER PORTION OF THE FILTER AND EXTENDING INTO THE IVC. DESPITE ATTEMPT AT ASPIRATION WITH CAT 8 DEVICE THE CLOT WAS ADHERENT TO THE FILTER. IT WAS DECIDED TO LEAVE THE FILTER IN PLACE.".
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2469245 | COOK CELECT® PLATINUM NAVALIGN JUGULAR VENA CAVA FILTER SET | DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR | DTK | WILLIAM COOK EUROPE | G34309 | E3643753 | 10827002343099 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Male | Life Threatening |