CHECK-IT
Report
- Report Number
- 3016521623-2022-00292
- Event Type
- Malfunction
- Date Received
- August 25, 2022
- Date of Event
- August 11, 2022
- Report Date
- August 18, 2022
- Manufacturer
- LUCIRA HEALTH, INC.
- Product Code
- QJR
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
Narratives
PRODUCT HAS NOT BEEN RETURNED. THEREFORE, A DHR REVIEW WAS COMPLETED FOR KIT LOT NUMBER K08A110804223M2 WITH 0 DISCREPANCIES FOUND (ALL LOTS PASSED IN TOTAL PER THE SAMPLING PLAN). A REVIEW OF EXISTING CAPAS, SCARS AND NCMRS WERE COMPLETED AND THERE ARE NO PRIOR RECORDS RELATED TO "FALSE POSITIVE" FAILURE MODE FOR THIS LOT. THERE ARE 0 ADDITIONAL COMPLAINTS FROM THIS CUSTOMER ASSOCIATED WITH A "FALSE POSITIVE" FAILURE MODE, PRIOR TO THE REPORTED RECEIPT DATE OF (B)(6) 2022. BASED ON REVIEW OF THE PRODUCT'S FMEAS AND RISK ASSESSMENT DOCUMENTATION, FALSE POSITIVE TEST RESULTS ARE A KNOWN POSSIBLE OUTCOME REGARDING THIS ISSUE UNDER EVALUATION, REFER TO FMEA-001 AND FMEA-004. THE COMPLAINT RATE FOR "FALSE POSITIVE" IS UNDER THE EXPECTED THRESHOLD OF 2% (LABEL CLAIM)/1% (INTERNAL WARNING LIMIT). LUCIRA HEALTH WILL CONTINUE TO MONITOR TRENDS RELATED TO FALSE POSITIVE RESULTS IN ACCORDANCE WITH POST-MARKET SURVEILLANCE PROCESS. DHR REVIEW KIT LOT# K08A110804223M2: SAMPLE VIAL LOT DHR REVIEW: 2202284, 2202439, 2203407, 2203138, 2203139, 2203158, 2203162, 2203222, 2203302, 2203362, 2203303 (ASSOCIATED INTERNAL LOT # 220124-1I, 220124-4E, 220124-4X, 220124-4V, 220124-4W, 220124-4P, 220124-7A, 220124-7E, 220124-7P, 220124-7X, 220124-7W) TEST LOT DHRS REVIEW: 2203616, 2203243, 2203243, 2203243, 2203245, 2203245, 2203245, 2203246, 2203305, 2203304, 2203304, 2203351, 2203351, 2203352, 2203352, 2203400, 2203351, 2203400, 2203401, 2203414 (ASSOCIATED INTERNAL LOT # 220305-30W, 220305-12C, 220305-12B, 220305-12A, 220305-12H, 220305-12G, 220305-12I, 220305-12K, 220305-15I, 220305-15E, 220305-15G, 220305-15W, 220305-15U, 220305-, 5Z, 220305-15Y, 220305-15S, 220305-15V, 220305-18L, 220305-18M, 220305-18R BASED ON THE LIMITED INFORMATION AVAILABLE, ROOT CAUSE CANNOT BE DETERMINED. POTENTIAL ROOT CAUSES INCLUDE BUT ARE NOT LIMITED TO: ENVIRONMENTAL CONTAMINATION, LOW VIRAL LOAD, AND/OR DEVICE FAILURE. A SUPPLEMENTAL REPORT WILL BE FILED IF ANY FURTHER INVESTIGATION AND/OR ADDITIONAL INFORMATION IS OBTAINED. BASED ON THE COMPLAINT REVIEW, NO HARM WAS REPORTED WITHIN THE COMPLAINT. THIS DEVICE IS MARKETED UNDER EUA (B)(4) CHECK-IT.
ONE DEVICE REPORTED AS HAVING AN ALLEGED FALSE POSITIVE RESULT. THE COMPLAINANT STATED THE LUCIRA TEST WAS PERFORMED WITH A POSITIVE RESULT. AN ACCULA PCR TEST WAS TAKEN AFTER THE POSITIVE RESULT WAS RECEIVED WITH A NEGATIVE RESULT FOR CONFIRMATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2393304 | CHECK-IT | COVID-19 TEST KIT | QJR | LUCIRA HEALTH, INC. | K08A110804223M2 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown | Other |