PVC DRAIN WITH TROCAR
Report
- Report Number
- 1018233-2022-06665
- Event Type
- Malfunction
- Date Received
- August 25, 2022
- Date of Event
- August 15, 2022
- Report Date
- November 7, 2022
- Manufacturer
- C.R. BARD, INC. (COVINGTON) -1018233
- Product Code
- GCY
- UDI-DI
- 00801741049385
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CO, US
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
THE INVESTIGATION IS STILL IN PROGRESS. ONCE THE INVESTIGATION IS COMPLETE A SUPPLEMENTAL REPORT WILL BE FILED.
UPON FURTHER REVIEW, BD HAS DETERMINED THAT THIS EVENT IS NOT REPORTABLE. H11: SECTION A THROUGH F - THE INFORMATION PROVIDED BY BD REPRESENTS ALL THE KNOWN INFORMATION AT THIS TIME. DESPITE GOOD FAITH EFFORTS TO OBTAIN ADDITIONAL INFORMATION, THE COMPLAINANT / REPORTER WAS UNABLE OR UNWILLING TO PROVIDE ANY FURTHER PATIENT, PRODUCT, OR PROCEDURAL DETAILS TO BD. H3 OTHER TEXT : THE DEVICE WAS NOT RETURNED.
IT WAS REPORTED THAT WOUND DRAIN MATERIAL# 0080720, MATERIAL# 0080710 AND MATERIAL# 0070310 DID NOT FIT WITH SILICONE EVACUATOR BULBS (MATERIAL# 0070740). CUSTOMER STATED THAT FIRST TWO DRAINS MATERIAL# 0080720, MATERIAL# 0080710 DID NOT COME WITH AN ADAPTOR. THE LAST ONE MATERIAL# 0070310 HAD AN ADAPTOR BUT DID NOT FIT WITH THE EVACUATOR BULB. PER FOLLOW-UP VIA EMAIL ON (B)(6) 2022 , BD REPRESENTATIVE SUGGESTED CUSTOMER THAT MATERIAL# 0080710 AND MATERIAL# 0080720 WERE PVC DRAINS AND ADVISED THAT THOSE WERE NOT CONNECTING BECAUSE THOSE WERE NOT LIKE MATERIALS, AS A PVC DRAIN CANNOT CONNECT TO THE SILICONE BULB. REPRESENTATIVE ADVISED THAT SILICONE DRAIN MATERIAL# 0070310 SHOULD CONNECT WITH SILICONE EVACUATOR MATERIAL# 0070740 SINCE THEY BOTH WERE SILICONE MATERIALS, AND THE CONNECTORS WERE WITHIN THE SILICONE DRAIN PACKAGING. PER ADDITIONAL INFORMATION RECEIVED ON 15-AUG-2022, CUSTOMER REPORTED THAT THEY WERE STILL HAVING ISSUES WITH DRAINS. THE BULBS THEY ORDERED WERE NOT THE GRENADE TYPE. PER ADDITIONAL INFORMATION RECEIVED ON 16-AUG-2022, CUSTOMER REPORTED THAT THEY WERE HAVING ISSUES WITH SOME HEMOVAC DRAINS. THEY USED JACKSON-PRATT (J/P) DRAINS MORE AND THOSE WERE THE ONES THEY WERE HAVING ISSUES WITH.
IT WAS REPORTED THAT WOUND DRAIN MATERIAL# 0080720, MATERIAL# 0080710 AND MATERIAL# 0070310 DID NOT FIT WITH SILICONE EVACUATOR BULBS (MATERIAL# 0070740). CUSTOMER STATED THAT FIRST TWO DRAINS MATERIAL# 0080720, MATERIAL# 0080710 DID NOT COME WITH AN ADAPTOR. THE LAST ONE MATERIAL# 0070310 HAD AN ADAPTOR BUT DID NOT FIT WITH THE EVACUATOR BULB. PER FOLLOW-UP VIA EMAIL ON 02-AUG-2022, BD REPRESENTATIVE SUGGESTED CUSTOMER THAT MATERIAL# 0080710 AND MATERIAL# 0080720 WERE PVC DRAINS AND ADVISED THAT THOSE WERE NOT CONNECTING BECAUSE THOSE WERE NOT LIKE MATERIALS, AS A PVC DRAIN CANNOT CONNECT TO THE SILICONE BULB. REPRESENTATIVE ADVISED THAT SILICONE DRAIN MATERIAL# 0070310 SHOULD CONNECT WITH SILICONE EVACUATOR MATERIAL# 0070740 SINCE THEY BOTH WERE SILICONE MATERIALS, AND THE CONNECTORS WERE WITHIN THE SILICONE DRAIN PACKAGING. PER ADDITIONAL INFORMATION RECEIVED ON 15-AUG-2022, CUSTOMER REPORTED THAT THEY WERE STILL HAVING ISSUES WITH DRAINS. THE BULBS THEY ORDERED WERE NOT THE GRENADE TYPE. PER ADDITIONAL INFORMATION RECEIVED ON 16-AUG-2022, CUSTOMER REPORTED THAT THEY WERE HAVING ISSUES WITH SOME HEMOVAC DRAINS. THEY USED JACKSON-PRATT (J/P) DRAINS MORE AND THOSE WERE THE ONES THEY WERE HAVING ISSUES WITH. PER FOLLOW-UP VIA EMAIL ON 14SEP2022, CUSTOMER HAD RESOLVED THE PROBLEM.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2681515 | PVC DRAIN WITH TROCAR | WOUND DRAIN | GCY | C.R. BARD, INC. (COVINGTON) -1018233 | 0043650 | UNK | 00801741049385 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown | Other |