FDA Adverse Event Malfunction Summary report: N

TPUC3

MDR report key: 15292741 · Received August 25, 2022

Report

Report Number
1823260-2022-02540
Event Type
Malfunction
Date Received
August 25, 2022
Date of Event
July 25, 2022
Report Date
October 24, 2022
Manufacturer
ROCHE DIAGNOSTICS
Product Code
JGQ
UDI-DI
04015630919536
PMA / PMN Number
K141925
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AR
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

PER PRODUCT LABELING, "URINE, CSF AND CONTROL SAMPLES WITH A PROTEIN CONCENTRATION ABOVE 7000 MG/L (700 MG/DL) MAY CLOG THE INSTRUMENT LINES." THE INVESTIGATION IS ONGOING.

Additional Manufacturer Narrative · 0

THE PATIENT HAD VIRAL MENINGITIS AT THE TIME OF SAMPLE COLLECTION. THE INVESTIGATION DID NOT IDENTIFY A PRODUCT PROBLEM. THE CAUSE OF THE EVENT COULD NOT BE DETERMINED.

Additional Manufacturer Narrative · 0

THE INFORMATION IN THE PREVIOUS SUPPLEMENTAL REPORT WAS INCORRECT. THE CORRECT INFORMATION IS: THE INVESTIGATION FOUND THAT THE ISSUE IS CONSISTENT WITH ELEVATED PROTEIN CONCENTRATION IN THE PATIENT'S SAMPLE. PER PRODUCT LABELING, "URINE, CSF AND CONTROL SAMPLES WITH A PROTEIN CONCENTRATION ABOVE 7000 MG/L (700 MG/DL) MAY CLOG THE INSTRUMENT LINES."

Description of Event or Problem · 0

THE INITIAL REPORTER RECEIVED QUESTIONABLE TPUC3 TOTAL PROTEIN CSF GEN.3 RESULTS ON A COBAS 4000 C311 ANALYZER WITH SERIAL NUMBER (B)(4). THE INITIAL CSF TP RESULT WAS 13.85 G/L. THE HIGH RESULT PROMPTED THE CUSTOMER TO SEND THE SAMPLE TO ANOTHER LAB TO REPEAT IT. THE REPEAT RESULT FROM THE OTHER LAB WITH THE SAME REAGENT LOT ON A C8000 ANALYZER WAS 14 G/L. THE SECOND LAB REFERRED THE SAMPLE TO A THIRD REFERENCE LAB FOR ELECTROPHORESIS AND PANDY SULFOSALICYLIC ACID QUALITATIVE TESTING AND THE RESULTS CAME BACK WITHIN NORMAL RANGE. THE ELECTROPHORESIS RESULT MATCHED THE PATIENT'S CLINICAL PICTURE. THE INITIAL RESULT WAS REPORTED OUTSIDE OF THE LABORATORY. THE ELECTROPHORESIS RESULT WAS DEEMED CORRECT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
432843 TPUC3 TURBIDIMETRIC, TOTAL PROTEIN JGQ ROCHE DIAGNOSTICS TPUC G3 59944801 04015630919536

Patients

Seq Age Sex Outcome Treatment
1 Unknown