FDA Adverse Event Malfunction Summary report: N

ICEROD CX 90 DEGREE NEEDLE/VL

MDR report key: 15289088 · Received August 24, 2022

Report

Report Number
2134265-2022-08656
Event Type
Malfunction
Date Received
August 24, 2022
Date of Event
May 5, 2022
Report Date
November 2, 2022
Manufacturer
BTG YOKNEAM
Product Code
GEH
UDI-DI
07290104830587
PMA / PMN Number
K121251
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

D3 - MANUFACTURER ADDRESS 1: (B)(6).

Additional Manufacturer Narrative · 0

MANUFACTURER ADDRESS 1: TAVOR BUILDING 1, INDUSTRIAL PARK. DEVICE EVAL BY MANUFACTURER: THE COMPLAINT NEEDLE WAS RETURNED FOR PRODUCT ANALYSIS. VISUAL ANALYSIS OF THE NEEDLE SHOWED NO OBVIOUS DAMAGE. THE NEEDLE WAS LEAK AND PERFORMANCE TESTED. AN AIR LEAK WAS SEEN FROM THE HANDLE-SHAFT CONNECTION. THE SHRINK WRAP WAS REMOVED FROM THE HANDLE AND MICROSCOPIC ANALYSIS WAS PERFORMED. MICROSCOPIC ANALYSIS OF THE HANDLE-SHAFT CONNECTION SHOWED A LACK OF GLUE.

Description of Event or Problem · 0

IT WAS REPORTED THE NEEDLE GAS LEAK OCCURRED. DURING PREPARATION FOR A RENAL CELL CARCINOMA (RCC) TREATMENT PROCEDURE, INTEGRITY TESTING OF AN ICEROD CX 90 DEGREE NEEDLE/VL CRYOABLATION NEEDLE REVEALED A GAS LEAK AT AN UNKNOWN LOCATION. THE NEEDLE WAS EXCHANGED AND THE PROCEDURE WAS COMPLETED WITH NO PATIENT COMPLICATIONS.

Description of Event or Problem · 0

IT WAS REPORTED THE NEEDLE GAS LEAK OCCURRED. DURING PREPARATION FOR A RENAL CELL CARCINOMA (RCC) TREATMENT PROCEDURE, INTEGRITY TESTING OF AN ICEROD CX 90 DEGREE NEEDLE/VL CRYOABLATION NEEDLE REVEALED A GAS LEAK AT AN UNKNOWN LOCATION. THE NEEDLE WAS EXCHANGED AND THE PROCEDURE WAS COMPLETED WITH NO PATIENT COMPLICATIONS. IT WAS FURTHER REPORTED THAT THE LEAK WAS COMING FROM THE GREY TUBE TO HANDLE JOINT.

Description of Event or Problem · 0

IT WAS REPORTED THE NEEDLE GAS LEAK OCCURRED. DURING PREPARATION FOR A RENAL CELL CARCINOMA (RCC) TREATMENT PROCEDURE, INTEGRITY TESTING OF AN ICEROD CX 90 DEGREE NEEDLE/VL CRYOABLATION NEEDLE REVEALED A GAS LEAK AT AN UNKNOWN LOCATION. THE NEEDLE WAS EXCHANGED AND THE PROCEDURE WAS COMPLETED WITH NO PATIENT COMPLICATIONS. IT WAS FURTHER REPORTED THAT THE LEAK WAS COMING FROM THE GREY TUBE TO HANDLE JOINT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
434320 ICEROD CX 90 DEGREE NEEDLE/VL UNIT, CRYOSURGICAL, ACCESSORIES GEH BTG YOKNEAM FPRPR3533 U3661 07290104830587

Patients

Seq Age Sex Outcome Treatment
1 Unknown