ICEROD CX 90 DEGREE NEEDLE/VL
Report
- Report Number
- 2134265-2022-08656
- Event Type
- Malfunction
- Date Received
- August 24, 2022
- Date of Event
- May 5, 2022
- Report Date
- November 2, 2022
- Manufacturer
- BTG YOKNEAM
- Product Code
- GEH
- UDI-DI
- 07290104830587
- PMA / PMN Number
- K121251
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
D3 - MANUFACTURER ADDRESS 1: (B)(6).
MANUFACTURER ADDRESS 1: TAVOR BUILDING 1, INDUSTRIAL PARK. DEVICE EVAL BY MANUFACTURER: THE COMPLAINT NEEDLE WAS RETURNED FOR PRODUCT ANALYSIS. VISUAL ANALYSIS OF THE NEEDLE SHOWED NO OBVIOUS DAMAGE. THE NEEDLE WAS LEAK AND PERFORMANCE TESTED. AN AIR LEAK WAS SEEN FROM THE HANDLE-SHAFT CONNECTION. THE SHRINK WRAP WAS REMOVED FROM THE HANDLE AND MICROSCOPIC ANALYSIS WAS PERFORMED. MICROSCOPIC ANALYSIS OF THE HANDLE-SHAFT CONNECTION SHOWED A LACK OF GLUE.
IT WAS REPORTED THE NEEDLE GAS LEAK OCCURRED. DURING PREPARATION FOR A RENAL CELL CARCINOMA (RCC) TREATMENT PROCEDURE, INTEGRITY TESTING OF AN ICEROD CX 90 DEGREE NEEDLE/VL CRYOABLATION NEEDLE REVEALED A GAS LEAK AT AN UNKNOWN LOCATION. THE NEEDLE WAS EXCHANGED AND THE PROCEDURE WAS COMPLETED WITH NO PATIENT COMPLICATIONS.
IT WAS REPORTED THE NEEDLE GAS LEAK OCCURRED. DURING PREPARATION FOR A RENAL CELL CARCINOMA (RCC) TREATMENT PROCEDURE, INTEGRITY TESTING OF AN ICEROD CX 90 DEGREE NEEDLE/VL CRYOABLATION NEEDLE REVEALED A GAS LEAK AT AN UNKNOWN LOCATION. THE NEEDLE WAS EXCHANGED AND THE PROCEDURE WAS COMPLETED WITH NO PATIENT COMPLICATIONS. IT WAS FURTHER REPORTED THAT THE LEAK WAS COMING FROM THE GREY TUBE TO HANDLE JOINT.
IT WAS REPORTED THE NEEDLE GAS LEAK OCCURRED. DURING PREPARATION FOR A RENAL CELL CARCINOMA (RCC) TREATMENT PROCEDURE, INTEGRITY TESTING OF AN ICEROD CX 90 DEGREE NEEDLE/VL CRYOABLATION NEEDLE REVEALED A GAS LEAK AT AN UNKNOWN LOCATION. THE NEEDLE WAS EXCHANGED AND THE PROCEDURE WAS COMPLETED WITH NO PATIENT COMPLICATIONS. IT WAS FURTHER REPORTED THAT THE LEAK WAS COMING FROM THE GREY TUBE TO HANDLE JOINT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 434320 | ICEROD CX 90 DEGREE NEEDLE/VL | UNIT, CRYOSURGICAL, ACCESSORIES | GEH | BTG YOKNEAM | FPRPR3533 | U3661 | 07290104830587 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |