MEDTRONIC TRANSCATHETER VALVE
Report
- Report Number
- 2025587-2022-02368
- Event Type
- Injury
- Date Received
- August 24, 2022
- Date of Event
- July 5, 2022
- Report Date
- August 24, 2022
- Manufacturer
- MEDTRONIC HEART VALVES DIVISION
- Product Code
- NPT
- PMA / PMN Number
- P130021
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NL
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
CITATION: GOLTSTEIN ET AL. REDUCTION OF GASTROINTESTINAL BLEEDING IN PATIENTS WITH HEYDE SYNDROME UNDERGOING TRANSCATHETER AORTIC VALVE IMPLANTATION. CIRC CARDIOVASC INTERV. 2022 JUL;15(7):E011848. DOI: 10.1161/CIRCINTERVENTIONS.122.011848. EPUB 2022 JUL 5. EARLIEST DATE OF PUBLICATION USED FOR DATE OF EVENT. MEDTRONIC PRODUCTS REFERENCED: COREVALVE AND EVOLUT R. EARLIEST APPROVED PRODUCT USED FOR PRODUCT CODE AND PMA#. NO UNIQUE DEVICE IDENTIFIER (SERIAL/LOT) NUMBERS WERE PROVIDED; WITHOUT THIS INFORMATION IT COULD NOT BE DETERMINED WHETHER THESE OBSERVATIONS HAVE BEEN PREVIOUSLY REPORTED. WITHOUT RETURN OF THE PRODUCT NO DEFINITIVE CONCLUSION CAN BE MADE REGARDING THE CLINICAL OBSERVATIONS. MEDTRONIC IS SUBMITTING THIS REPORT TO COMPLY WITH FDA REPORTING REGULATIONS UNDER 21 CFR PARTS 4 AND 803. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION AND HAS PROVIDED AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY ¿DEFECTS¿ OR HAS ¿MALFUNCTIONED¿. THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REGULATORY REPORTING. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE THEM BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
MEDTRONIC RECEIVED INFORMATION VIA LITERATURE REGARDING REDUCTION OF GASTROINTESTINAL BLEEDING IN PATIENTS WITH HEYDE SYNDROME UNDERGOING TRANSCATHETER AORTIC VALVE IMPLANTATION (TAVI). ALL DATA WERE COLLECTED FROM A SINGLE CENTER BETWEEN DECEMBER 2008 AND JUNE 2020. THE STUDY POPULATION INCLUDED 280 PATIENTS WHO WERE PREDOMINANTLY MALE WITH A MEAN AGE OF 77 YEARS. MULTIPLE MANUFACTURER¿S DEVICES WERE IMPLANTED IN THE STUDY POPULATION; 210 PATIENTS WERE IMPLANTED WITH A MEDTRONIC COREVALVE (N=94) OR EVOLUT R (N=116) BIOPROSTHETIC VALVE. NO UNIQUE DEVICE IDENTIFIER NUMBERS WERE PROVIDED. AMONG ALL PATIENTS, 25 DEATHS OCCURRED WITHIN ONE YEAR. NO FURTHER DETAILS WERE PROVIDED ON THE DEATHS. THERE WAS NO STATEMENT OF CAUSAL OR CONTRIBUTORY RELATIONSHIP BETWEEN MEDTRONIC PRODUCT AND THE DEATHS. AMONG ALL PATIENTS, ADVERSE EVENTS INCLUDED: BLEEDING EVENTS REQUIRING INTERVENTION, AND MILD TO MODERATE PARAVALVULAR LEAK (PVL). BASED ON THE AVAILABLE INFORMATION MEDTRONIC PRODUCT MAY HAVE BEEN ASSOCIATED WITH THE ADVERSE EVENTS. NO ADDITIONAL ADVERSE PATIENT EFFECTS OR PRODUCT PERFORMANCE ISSUES WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 478675 | MEDTRONIC TRANSCATHETER VALVE | AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIV | NPT | MEDTRONIC HEART VALVES DIVISION | MDT-TRANS VALVE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 77 YR | Male | Required Intervention| L |