FDA Adverse Event Malfunction Summary report: N

BINAXNOW COVID-19 ANTIGEN SELF-TEST

MDR report key: 15283117 · Received August 24, 2022

Report

Report Number
1221359-2022-04320
Event Type
Malfunction
Date Received
August 24, 2022
Date of Event
August 10, 2022
Report Date
August 30, 2022
Manufacturer
ABBOTT DIAGNOSTICS SCARBOROUGH, INC.
Product Code
QKP
UDI-DI
00811877011408
PMA / PMN Number
EUA210264
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

THE REMAINDER OF THE INVESTIGATION REMAINS IN PROGRESS. A SUPPLEMENTAL REPORT WILL BE PROVIDED AFTER COMPLETION.

Additional Manufacturer Narrative · 0

CORRECTION : CORRECT DATE OF SUBMISSION OF INITIAL REPORT IS 24AUG2022. ADDITIONAL INFORMATION : TESTING WAS PERFORMED AT ABBOTT DIAGNOSTICS SCARBOROUGH, INC. ON RETAINED KIT LOT 210226 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR TEST KIT PART NUMBER 195-160 / LOT 210226 AND TEST BASE PART NUMBER 195-430WL / LOT 207462. THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. A REVIEW OF THE COMPLAINTS REPORTED AS FALSE POSITIVE PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) RELATED TO KIT LOT 210226 SHOWED THAT THE COMPLAINT RATE IS 0.000688%. IN CONCLUSION, ABBOTT DIAGNOSTICS SCARBOROUGH WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE; HOWEVER, IT COULD POSSIBLY BE RELATED TO ISSUES INCLUDING THE SELF-TEST USER PERFORMANCE, INTERPRETATION OF THE RESULT, OR THE SPECIFIC PATIENT SAMPLE.

Description of Event or Problem · 0

THE CONSUMER REPORTED TWO (2) FALSE POSITIVE RESULTS WITH BINAXNOW COVID-19 ANTIGEN SELF TEST PERFORMED ON AN (B)(6) 2022 WITH AN UNKNOWN SAMPLE. THIS REPORT IS FOR TEST ONE (1) OF TWO (2). TESTING WAS PERFORMED ON (B)(6) 2022 USING AN UNKNOWN PRODUCT THAT GENERATED A NEGATIVE RESULT. THE PATIENT BEGAN SHOWING SYMPTOMS ON (B)(6) 2022 WITH A FEVER AND COUGH. NO ADDITIONAL PATIENT INFORMATION, INCLUDING TREATMENT AND OUTCOME, WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
436792 BINAXNOW COVID-19 ANTIGEN SELF-TEST CORONAVIRUS ANTIGEN DETECTION TEST SYSTEM. QKP ABBOTT DIAGNOSTICS SCARBOROUGH, INC. 210226 00811877011408

Patients

Seq Age Sex Outcome Treatment
1 1 YR Male