FDA Adverse Event Malfunction Summary report: N

VITROS IMMUNODIAGNOSTIC PRODUCTS TOTAL B HCG II REAGENT PACK

MDR report key: 15282131 · Received August 24, 2022

Report

Report Number
3007111389-2022-00082
Event Type
Malfunction
Date Received
August 24, 2022
Date of Event
July 11, 2022
Report Date
August 24, 2022
Manufacturer
ORTHO-CLINICAL DIAGNOSTICS
Product Code
DHA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

THE INVESTIGATION DETERMINED THAT A NON-REPRODUCIBLE, DISCORDANT, HIGHER THAN EXPECTED VITROS HCG II (B-HCG) RESULT WAS OBTAINED FROM A SINGLE PATIENT SAMPLE WHEN TESTED USING VITROS B-HCG LOT 3390 ON A VITROS ECI IMMUNODIAGNOSTIC SYSTEM. A DEFINITIVE CAUSE OF THE EVENT WAS NOT ESTABLISHED. A VITROS B-HCG LOT 3390 REAGENT ISSUE CANNOT BE RULED OUT AS A CONTRIBUTOR TO THE EVENT. QUALITY CONTROL RESULTS LEADING UP TO THE EVENT INDICATE ACCEPTABLE REAGENT PERFORMANCE WHEN COMPARED TO THE CUSTOMERS ESTABLISHED BASELINE MEAN/ SD. HOWEVER, THE CUSTOMER DID NOT PROVIDE ANY MANUFACTURER LOT INFORMATION, THEREFORE PEER MEAN/ SD AND PACKAGE INSERT MEAN/ SD INFORMATION WAS UNABLE TO BE RETRIEVED FOR COMPARISON TO SUPPORT ACCEPTABLE PERFORMANCE. HOWEVER, CONTINUAL TRACKING AND TRENDING OF COMPLAINT DATA HAS NOT IDENTIFIED ANY SIGNALS TO SUGGEST THERE IS A SYSTEMIC QUALITY ISSUE WITH VITROS B-HCG REAGENT LOT 3390. AN INSTRUMENT ISSUE CANNOT BE RULED OUT AS A CONTRIBUTOR TO THE EVENT. AN ORTHO FIELD ENGINEER PERFORMED SERVICE ACTIONS OVER SEVERAL DAYS AT THE CUSTOMER SITE, WHICH INCLUDED PREVENTATIVE CLEANING OF THE SYSTEM AND SAMPLE ARM ADJUSTMENTS, REPLACING THE WELL WASH ASPIRATE FILTER, REPLACING THE WELL WASH NOZZLE, SERVICING THE SIGNAL REAGENT MODULE, SERVICING THE MICROWELL INCUBATOR, AND SERVICING THE REAGENT SUPPLY. HOWEVER, DIAGNOSTIC PRECISION TESTING WAS NOT CONDUCTED BEFORE OR AFTER THE SERVICE ACTIONS, AND NO QUALITY CONTROL DATA WAS PROVIDED FOLLOWING SERVICE. THEREFORE, AN INSTRUMENT ISSUE CANNOT BE COMPLETELY CONFIRMED OR RULED OUT AS A CONTRIBUTOR TO THE EVENT. IN ADDITION, PRE-ANALYTICAL SAMPLE PROCESSING COULD NOT BE RULED OUT AS A CONTRIBUTING FACTOR AS IT IS UNKNOWN IF THE CUSTOMER WAS FOLLOWING THE SAMPLE COLLECTION DEVICE MANUFACTURE¿S RECOMMENDATION FOR SAMPLE CENTRIFUGATION. IMPROPER PRE-ANALYTICAL SAMPLE HANDLING COULD HAVE CONTRIBUTED TO THIS EVENT. IT IS POSSIBLE THAT CELLULAR DEBRIS, DUE TO POOR SAMPLE PREPARATION, WAS PRESENT IN THE AFFECTED SAMPLE, ALTHOUGH THIS COULD NOT BE CONFIRMED. AS THE REPEAT VITROS B-HCG RESULT (<2.39 MIU/ML ) WAS DIFFERENT TO THE INITIAL VITROS B-HCG RESULT (19.87 MIU/ML) FOR PATIENT 1, PATIENT SAMPLE MIX UP CANNOT BE RULED OUT AS A CONTRIBUTOR TO THE EVENT. EMAIL ADDRESS FOR CONTACT OFFICE IS (B)(4).

Description of Event or Problem · 0

A CUSTOMER CONTACTED THE ORTHO CLINICAL DIAGNOSTICS (ORTHO) TECHNICAL SOLUTION CENTER (TSC) TO REPORT A NON-REPRODUCIBLE, DISCORDANT, HIGHER THAN EXPECTED VITROS HCG II (B-HCG) RESULT WAS OBTAINED FROM A SINGLE PATIENT SAMPLE WHEN TESTED USING VITROS B-HCG LOT 3390 ON A VITROS ECI IMMUNODIAGNOSTIC SYSTEM. PATIENT 1 VITROS B-HCG RESULT OF 19.87 MIU/ML VERSUS THE EXPECTED RESULT OF < 2.39 MIU/ML BIASED RESULTS OF THE MAGNITUDE AND DIRECTION OBSERVED MAY LEAD TO INAPPROPRIATE PHYSICIAN ACTION IF THEY WERE TO OCCUR UNDETECTED. THE HIGHER THAN EXPECTED VITROS B-HCG RESULT WAS NOT REPORTED FROM THE LABORATORY. NO TREATMENT WAS ALTERED, INITIATED OR STOPPED BASED ON THE INITIALLY REPORTED RESULT. ORTHO HAS NOT BEEN MADE AWARE OF ANY ALLEGATION OF PATIENT HARM AS A RESULT OF THIS EVENT. THIS REPORT CORRESPONDS TO ORTHO CLINICAL DIAGNOSTICS INC (ORTHO) COMPLAINT NUMBER (B)(4) AND IVD (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1827986 VITROS IMMUNODIAGNOSTIC PRODUCTS TOTAL B HCG II REAGENT PACK IN-VITRO DIAGNOSTICS DHA ORTHO-CLINICAL DIAGNOSTICS 3390

Patients

Seq Age Sex Outcome Treatment