I-STAT ACT CELITE CARTRIDGE
Report
- Report Number
- 2245578-2022-00114
- Event Type
- Malfunction
- Date Received
- August 23, 2022
- Date of Event
- August 12, 2022
- Report Date
- October 25, 2022
- Manufacturer
- ABBOTT POINT OF CARE
- Product Code
- JBP
- UDI-DI
- 10054749000156
- PMA / PMN Number
- K023582
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SN
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
APOC INCIDENT # (B)(4). APOC LABELING WILL BE EVALUATED DURING THE INVESTIGATION AS PERTAINING TO THE EVENT.
APOC INCIDENT: # (B)(4). THE INVESTIGATION WAS COMPLETED ON 20-OCT-2022. A REVIEW OF THE DEVICE HISTORY RECORD (DHR) CONFIRMED THE CARTRIDGE LOT PASSED RELEASE SPECIFICATIONS. RETAINED CARTRIDGE TESTING MET THE ACCEPTANCE CRITERIA IN APPENDIX 1 OF Q04.01.003 REV. AK (PRODUCT COMPLAINT LEVEL 2 AND LEVEL 3 INVESTIGATION PROCEDURE). NO DEFICIENCY HAS BEEN IDENTIFIED.
ON 12-AUG-2022, ABBOTT POINT OF CARE WAS CONTACTED BY A CUSTOMER REGARDING ACT KAOLIN CARTRIDGES THAT YIELDED DISCREPANT RESULTS ON A PATIENT. THERE WAS NO PATIENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT. RUN, TIME, RESULT. 1ST RUN, 1504H, 250 SECONDS. 2ND RUN, 1552H , >1000 SECONDS. 3RD RUN, 1559H, 181 SECONDS. PER I-STAT SYSTEM MANUAL (ART: 770715-00A), THE I-STAT KAOLIN ACTIVATED CLOTTING TIME (KAOLINACT) TEST IS AN IN VITRO DIAGNOSTIC TEST THAT USES FRESH, WHOLE BLOOD, AND IS USED TO MONITOR HIGH-DOSE HEPARIN ANTICOAGULATION FREQUENTLY ASSOCIATED WITH CARDIOVASCULAR SURGERY. AT THIS TIME THERE IS NO REASON TO SUSPECT A MALFUNCTION EXISTS. THE REPORTING DECISION WAS BASED ON THE LIMITED INFORMATION AVAILABLE THAT SUGGESTS THE PRODUCT WAS NOT PERFORMING WITHIN THE VARIABILITY OF THE ASSAY. AN INVESTIGATION IS UNDERWAY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2822373 | I-STAT ACT CELITE CARTRIDGE | ACT CELITE CARTRIDGE | JBP | ABBOTT POINT OF CARE | NA | R22131 | 10054749000156 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |