FDA Adverse Event Injury Summary report: N

IMPL TAPERED SCR-V SBM 3.7MM 3.5MM 10MM

MDR report key: 15271719 · Received August 22, 2022

Report

Report Number
0002023141-2022-02156
Event Type
Injury
Date Received
August 22, 2022
Date of Event
July 29, 2022
Report Date
February 2, 2023
Manufacturer
ZIMMER DENTAL
Product Code
DZE
UDI-DI
00889024019508
PMA / PMN Number
K061410
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

MULTIPLE MDR REPORTS WERE FILED FOR THIS EVENT. PLEASE SEE ASSOCIATED REPORT: 0002023141-2022-02155. ZIMMERBIOMET COMPLAINT NUMBER: 2124215. PMA/510(K) NUMBERS: K011028, K01322. A SUMMARY INVESTIGATION HAS BEEN COMPLETED FOR INFECTION EVENTS IDENTIFYING THAT A DEFINITIVE ROOT CAUSE CANNOT BE IDENTIFIED DUE TO THE WIDE RANGE OF EXTERNAL (NON-DESIGN/ NON-MANUFACTURING RELATED) FACTORS POTENTIALLY IMPACTING THE STERILITY OF THE IMPLANT AND ITS ENVIRONMENT. SHOULD ADDITIONAL INFORMATION BE RECEIVED WHICH INDICATES THAT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE INFECTION EVENT, AN ADDITIONAL REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 0

MULTIPLE MDR REPORTS WERE FILED FOR THIS EVENT. PLEASE SEE ASSOCIATED REPORT: 0002023141-2022-02155-1. THIS REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION AND DEVICE EVALUATION. THE FOLLOWING SECTIONS ARE BEING REPORTED: H6: TYPE OF INVESTIGATION CODES WERE UPDATED: 3331. DHR REVIEW WAS COMPLETED FOR THE SUBJECT LOT NUMBER 1248968. IT WAS CONFIRMED THAT ALL OPERATIONS AND INSPECTIONS WERE EXECUTED AS PER APPLICABLE PROCEDURE. NO DEVIATIONS OR NON-CONFORMANCES, WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT WAS NOTED AS PART OF THE DHR. LOT WAS INSPECTED AND PASSED ALL ACCEPTANCE CRITERIA BY QA. STERILIZATION RECORD OP# 150 WAS REVIEWED AND VERIFIED TO HAVE PASSED ALL STERILIZATION ACTIVITIES WITH NO ISSUES OR NONCONFORMITIES IDENTIFIED. COMPLAINT HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT NUMBER 1248968 FOR SIMILAR EVENTS AND NO OTHER COMPLAINT WAS IDENTIFIED. AS DOCUMENTED IN THE SUMMARY INVESTIGATION, CONTRIBUTING FACTORS FOR THESE REPORTED EVENTS LIKELY EXIST OUTSIDE OF ZIMMER BIOMET CONTROL, INCLUDING THOSE RELATED TO PATIENT BIOLOGICAL FACTORS/CONDITION AND SURGICAL TECHNIQUE. BASED ON THE SUMMARY INVESTIGATION, NO MALFUNCTION OCCURRED UPON INVESTIGATION. THE REPORTED EVENT REMAINS NON-VERIFIABLE AS IT IS A MEDICAL CONDITION.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE IMPLANTS ON TEETH SITES #23 AND #24 WERE REMOVED DUE TO INFECTION.

Description of Event or Problem · 0

NO ADDITIONAL EVENT INFORMATION WAS RECEIVED AT THE TIME OF THIS REPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2389890 IMPL TAPERED SCR-V SBM 3.7MM 3.5MM 10MM DENTAL IMPLANT DZE ZIMMER DENTAL TSVB10 1248968 00889024019508

Patients

Seq Age Sex Outcome Treatment
1 62 YR Male Required Intervention