FDA Adverse Event Injury Summary report: N

ENROUTE NEUROPROTECTION SYSTEM

MDR report key: 15266846 · Received August 22, 2022

Report

Report Number
3014526664-2022-00124
Event Type
Injury
Date Received
August 22, 2022
Date of Event
July 19, 2022
Report Date
August 22, 2022
Manufacturer
SILK ROAD MEDICAL INC.
Product Code
NTE
UDI-DI
00811311020829
PMA / PMN Number
K153485
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE NEUROPROTECTION SYSTEM (NPS) ASSOCIATED WITH THIS COMPLAINT WAS NOT RETURNED TO THE MANUFACTURER FOR ANALYSIS. A REVIEW OF THE MANUFACTURING RECORDS FOR THIS DEVICE WAS COMPLETED AND NO ISSUES WERE IDENTIFIED THAT COULD HAVE LED TO THE ADVERSE EVENT REPORTED. THE DEVICE RESPONSIBLE FOR THE DISSECTION WAS NOT IDENTIFIED BY THE USER. AT THIS TIME, IT IS UNKNOWN IF THE REPORTED FAILURE IS RELATED TO THE MICROPUNCTURE KIT (MPK) OR THE ENROUTE NPS; HENCE, THE EVENT WILL BE REPORTED OUT OF ABUNDANCE OF CAUTION. A FOLLOW-UP MDR WILL BE SUBMITTED IF ADDITIONAL INFORMATION IS OBTAINED. COMPLAINTS WILL CONTINUE TO BE REVIEWED AND MONITORED FOR TRENDS.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING A TRANSCAROTID ARTERY REVASCULARIZATION (TCAR) PROCEDURE, A DISSECTION WAS NOTED ON THE INITIAL ANGIOGRAM FOLLOWING ENROUTE ARTERIAL SHEATH INSERTION. PER THE PHYSICIAN, A DISSECTION WAS SUSPECTED AFTER USING MICROPUNCTURE KIT (MPK), WHICH BECAME MUCH WORSE AND MORE NOTICEABLE AFTER ARTERIAL SHEATH PLACEMENT. THE PHYSICIAN ELECTED CONVERT TO CAROTID ENDARTERECTOMY (CEA). THE DEVICE RESPONSIBLE FOR THE DISSECTION WAS NOT IDENTIFIED BY THE USER. AT THIS TIME, IT IS UNKNOWN IF THE REPORTED FAILURE IS RELATED TO THE MPK OR THE ENROUTE NEUROPROTECTION SYSTEM (NPS); HENCE, THE EVENT WILL BE REPORTED OUT OF ABUNDANCE OF CAUTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2390364 ENROUTE NEUROPROTECTION SYSTEM TEMPORARY CAROTID CATHETER FOR EMBOLIC CAPTURE NTE SILK ROAD MEDICAL INC. SR-200-NPS 302166 00811311020829

Patients

Seq Age Sex Outcome Treatment
1 78 YR Female Required Intervention