CODA BALLOON CATHETER
Report
- Report Number
- 1820334-2009-00631
- Event Type
- Injury
- Date Received
- November 10, 2009
- Date of Event
- January 31, 2007
- Report Date
- October 15, 2009
- Manufacturer
- COOK INC
- Product Code
- GBZ
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- PHYSICIAN
Narratives
EACH CODA BALLOON IS SHIPPED WITH AN IFU LISTING WARNINGS, PRECAUTIONS, AND INSTRUCTIONS FOR USE. SPECIFICALLY, IT WARNS THAT OVER INFLATION MAY RESULT IN DAMAGE AND/OR VESSEL RUPTURE AND THAT THE DEVICE "IS INTENDED FOR TEMPORARY OCCLUSION OF LARGE VESSELS, OR TO EXPAND VASCULAR PROSTHESES." IN THE WARNINGS SECTION, IT IS STATED THAT THE "40MM BALLOON CATHETER SHOULD NOT BE USED IN THE VESSELS LESS THAN 24MM IN DIAMETER." THE INFO THAT WAS RETURNED WAS LIMITED. THE RPN AND LOT NUMBER OF THE DEVICE ARE UNK, MAKING A SEARCH OF OUR MFG RECORDS DIFFICULT. THE COMPLAINT INFO WAS RECEIVED. THE REP INDICATED THAT THE PROCEDURE TOOK PLACE IN 2007 AND THAT A CODA BALLOON WAS BEING USED FOR ANGIOPLASTY OF SEVERE ATHEROSCLEROTIC PLAQUE. WE WERE UNABLE TO CONFIRM THE DATE IN WHICH THE CODA BALLOON WAS USED. THE USE OF THE BALLOON FOR ANGIOPLASTY AND THE DISEASED CONDITION OF THE PT'S ILIAC ARTERY COULD HAVE CONTRIBUTED TO VESSEL DISSECTION. OVER-INFLATION OF THE BALLOON CAN RESULT IN INJURY TO THE VESSEL WALL AND/OR RUPTURE MAY OCCUR. WITH THE INFO PROVIDED AND DESCRIPTION OF THE EVENT IT IS DIFFICULT TO DETERMINE A ROOT CAUSE. THE INTENDED USE OF THE DEVICE AND THE APPROPRIATE WARNINGS ARE DOCUMENTED IN THE SUPPLIED IFU. WE WILL NOTIFY THE APPROPRIATE PERSONNEL OF THE EVENT AND CONTINUE TO MONITOR FOR SIMILAR COMPLAINTS.
A MALE PT UNDERWENT AAA REPAIR IN 2007. THE PT RECEIVED ANOTHER MFR'S ENDOGRAFTS. DURING THE PROCEDURE, SHEATH INSERTION (ANOTHER MFR'S) AND ANGIOPLASTY, WITH A COOK CODA BALLOON, CAUSED DISLODGEMENT OF UNDERLYING SEVERE ATHEROSCLEROTIC PLAQUE, CAUSING DISSECTION IN THE RIGHT COMMON AND EXTERNAL ILIAC VESSELS. THE DISSECTION WAS REPAIRED USING BALLOON EXPANDABLE STENTS (ANOTHER MFR'S). THE PT TOLERATED THE PROCEDURE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CODA BALLOON CATHETER | GBZ CATHETER, CHOLANGIOGRAPHY | GBZ | COOK INC | NA | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 70 YR | Required Intervention |