QUICK SET PARADIGM
Report
- Report Number
- 8021545-2009-00041
- Event Type
- Malfunction
- Date Received
- November 2, 2009
- Date of Event
- July 9, 2009
- Report Date
- October 29, 2009
- Manufacturer
- UNOMEDICAL A/S
- Product Code
- FPA
- PMA / PMN Number
- K991759
- Removal / Correction Number
- Z-1705-2009
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- NOT APPLICABLE
Narratives
EVAL SUMMARY: ONE USED DEVICE WAS RETURNED FOR EVAL. USED DEVICES: THE TUBING WAS VISUALLY INSPECTED FOR ANY TEARS OR CRACKS ON THE TUBING ITSELF AND ON THE ATTACHED CONNECTOR PARTS. FURTHERMORE A FLOW TEST AND A P-CAP CONNECTOR VENT TEST WERE PERFORMED. THE MEMBRANE IN THE P-CAP CONNECTOR WAS HUMID AND THE SAMPLE ALSO FAILED THE P-CAP CONNECTOR VENT TEST. FURTHERMORE THE USED SAMPLE FAILED FLOW TEST OF THE TUBING. UNUSED DEVICES: NO UNUSED SAMPLES WERE RETURNED FOR EVAL. REFERENCE SAMPLES: THE REFERENCE MATERIAL WAS TESTED AND ALL SAMPLES WERE FOUND TO BE WITHIN SPECS. A REVIEW OF THE RELEVANT DEVICE HISTORY RECORDS AND A SEARCH FOR RELEVANT DEVIATIONS OR SIMILAR COMPLAINTS RELATED TO THE SPECIFIC DEVICE RESULTED IN NO SIMILAR COMPLAINTS FOR THE INVOLVED LOT NUMBER 8200752. NO RELEVANT DEVIATIONS DURING MFG WERE RECORDED.
PT STATES THAT INSULIN IN POURING OUT OF THE INFUSION SET TUBING WITHOUT PUMP RECORDING THE FLOW. PT HAS HAD PUMP REPLACED LAST MONTH FOR THE SAME ISSUE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | QUICK SET PARADIGM | FPA | UNOMEDICAL A/S | MMT-399 | 8200752 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |