FDA Adverse Event Malfunction Summary report: N

VS3 IRIDIUM - LIGHT INTEGRATOR (BEAMCOMBINER)

MDR report key: 15263557 · Received August 21, 2022

Report

Report Number
9616637-2022-00002
Event Type
Malfunction
Date Received
August 21, 2022
Date of Event
September 30, 2021
Report Date
August 21, 2022
Manufacturer
VISIONSENSE LTD.
Product Code
OWN
PMA / PMN Number
K150018
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

THE DEVICE WAS RECEIVED FOR EVALUATION AND MEDTRONIC CONDUCTED AN INVESTIGATION BASED UPON ALL RECEIVED INFORMATION. THE REPORTED CONDITION WAS CONFIRMED. THE MOST LIKELY ROOT CAUSE WAS IDENTIFIED TO BE A DAMAGED FIBER CABLE CAUSED BY IMPROPER HANDLING. IMPROVEMENTS HAVE BEEN INITIATED TO MITIGATE THIS CONDITION. TO ENSURE PRODUCT OVERSIGHT, THIS COMPLAINT REPORT IS INCORPORATED INTO MEDTRONIC¿S COMPLAINT MONITORING AND TRACKING SYSTEM. THE MANUFACTURING RECORDS FOR EACH DEVICE ARE THOROUGHLY REVIEWED TO ENSURE THE PRODUCT MEETS ITS QUALITY SPECIFICATIONS. SHOULD NEW INFORMATION BECOME AVAILABLE THE FILE WILL BE RE-OPENED AND THE INVESTIGATION SUMMARY WILL BE AMENDED AS APPROPRIATE. MEDTRONIC IS SUBMITTING THIS REPORT TO COMPLY WITH FDA REPORTING REGULATIONS UNDER 21 CFR PARTS 4 AND 803. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION AND HAS PROVIDED AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY ¿DEFECTS¿ OR HAS ¿MALFUNCTIONED¿. THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REGULATORY REPORTING. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE THEM BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 0

DURING SERVICING, THE UNIT HAD A LASER ERROR AND THE SERVICE TECHNICIAN SMELLED SMOKE. THERE WAS NO PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1942129 VS3 IRIDIUM - LIGHT INTEGRATOR (BEAMCOMBINER) CONFOCAL OPTICAL IMAGING OWN VISIONSENSE LTD. 161-0001

Patients

Seq Age Sex Outcome Treatment
1 Unknown