BD VACUTAINER® SERUM BLOOD COLLECTION TUBE
Report
- Report Number
- 1024879-2022-00474
- Event Type
- Malfunction
- Date Received
- August 19, 2022
- Date of Event
- August 11, 2022
- Report Date
- August 26, 2022
- Manufacturer
- BECTON, DICKINSON & CO., (BD)
- Product Code
- JKA
- UDI-DI
- 50382903678155
- PMA / PMN Number
- BK050036
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
H.6. INVESTIGATION SUMMARY: MATERIAL #: 3678215; LOT/BATCH #: 2123154. BD HAD NOT RECEIVED SAMPLES OR PHOTOS FOR INVESTIGATION. THEREFORE, 30 RETENTION SAMPLES FROM BD INVENTORY WERE EVALUATED BY VISUAL EXAMINATION AND NO ISSUES WERE OBSERVED RELATING TO MISSING ADDITIVE AS ALL SAMPLES MET SPECIFICATIONS. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THERE WERE NO RELATED QUALITY ISSUES DURING MANUFACTURING OF THE PRODUCT. THIS COMPLAINT IS UNABLE TO BE CONFIRMED FOR THE INDICATED FAILURE MODE MISSING ADDITIVE. BD WAS NOT ABLE TO IDENTIFY A ROOT CAUSE FOR THE INDICATED FAILURE MODE. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS. H3 OTHER TEXT : SEE H.10.
IT WAS REPORTED WHEN USING THE BD VACUTAINER® SERUM BLOOD COLLECTION TUBE THERE WAS MISSING ADDITIVE THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER. THE CUSTOMER STATED: THERE WAS "NO SERUM IN THE TUBES."
IT WAS REPORTED WHEN USING THE BD VACUTAINER® SERUM BLOOD COLLECTION TUBE THERE WAS MISSING ADDITIVE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER. THE CUSTOMER STATED: THERE WAS "NO SERUM IN THE TUBES."
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2906318 | BD VACUTAINER® SERUM BLOOD COLLECTION TUBE | BLOOD SPECIMEN COLLECTION DEVICE | JKA | BECTON, DICKINSON & CO., (BD) | 367815 | 2123154 | 50382903678155 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |