FDA Adverse Event Malfunction Summary report: N

BD VACUTAINER® SERUM BLOOD COLLECTION TUBE

MDR report key: 15259024 · Received August 19, 2022

Report

Report Number
1024879-2022-00474
Event Type
Malfunction
Date Received
August 19, 2022
Date of Event
August 11, 2022
Report Date
August 26, 2022
Manufacturer
BECTON, DICKINSON & CO., (BD)
Product Code
JKA
UDI-DI
50382903678155
PMA / PMN Number
BK050036
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION SUMMARY: MATERIAL #: 3678215; LOT/BATCH #: 2123154. BD HAD NOT RECEIVED SAMPLES OR PHOTOS FOR INVESTIGATION. THEREFORE, 30 RETENTION SAMPLES FROM BD INVENTORY WERE EVALUATED BY VISUAL EXAMINATION AND NO ISSUES WERE OBSERVED RELATING TO MISSING ADDITIVE AS ALL SAMPLES MET SPECIFICATIONS. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THERE WERE NO RELATED QUALITY ISSUES DURING MANUFACTURING OF THE PRODUCT. THIS COMPLAINT IS UNABLE TO BE CONFIRMED FOR THE INDICATED FAILURE MODE MISSING ADDITIVE. BD WAS NOT ABLE TO IDENTIFY A ROOT CAUSE FOR THE INDICATED FAILURE MODE. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS. H3 OTHER TEXT : SEE H.10.

Description of Event or Problem · 0

IT WAS REPORTED WHEN USING THE BD VACUTAINER® SERUM BLOOD COLLECTION TUBE THERE WAS MISSING ADDITIVE THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER. THE CUSTOMER STATED: THERE WAS "NO SERUM IN THE TUBES."

Description of Event or Problem · 0

IT WAS REPORTED WHEN USING THE BD VACUTAINER® SERUM BLOOD COLLECTION TUBE THERE WAS MISSING ADDITIVE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER. THE CUSTOMER STATED: THERE WAS "NO SERUM IN THE TUBES."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2906318 BD VACUTAINER® SERUM BLOOD COLLECTION TUBE BLOOD SPECIMEN COLLECTION DEVICE JKA BECTON, DICKINSON & CO., (BD) 367815 2123154 50382903678155

Patients

Seq Age Sex Outcome Treatment
1 Unknown