T:SLIM X2 INSULIN PUMP WITH CONTROL-IQ TECHNOLOGY
Report
- Report Number
- 3013756811-2022-89044
- Event Type
- Malfunction
- Date Received
- August 19, 2022
- Date of Event
- July 28, 2022
- Report Date
- July 29, 2022
- Manufacturer
- TANDEM DIABETES CARE
- Product Code
- QFG
- UDI-DI
- 00850006613762
- PMA / PMN Number
- K201214
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
Narratives
NO PRODUCT WAS RETURNED FOR EVALUATION. SHOULD NEW RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PER TANDEM USER GUIDE: ONLY USE U-100 HUMALOG OR U-100 NOVOLOG WITH YOUR PUMP. ONLY U-100 HUMALOG AND NOVOLOG HAVE BEEN TESTED AND FOUND TO BE COMPATIBLE FOR USE IN THE PUMP. USE OF INSULIN WITH LESSER OR GREATER CONCENTRATION CAN RESULT IN UNDER DELIVERY OR OVER DELIVERY OF INSULIN. THIS CAN CAUSE HYPOGLYCEMIA (LOW BG) OR HYPERGLYCEMIA (HIGH BG) EVENTS.
IT WAS REPORTED THAT INTERMITTENT OCCLUSION ALARMS OCCURRED. THE CUSTOMER'S BLOOD GLUCOSE LEVEL RANGED FROM 230-359 MG/DL. DURING A SYSTEMS CHECK WITH TANDEM TECHNICAL SUPPORT (CTS) IT WAS DISCOVERED THAT APIDRA INSULIN WAS USED WITHIN THE CARTRIDGE. CTS EDUCATED CUSTOMER ON INSULIN/CARTRIDGE LABELING. REPORTEDLY, THE CUSTOMER CHANGED PUMP SUPPLIES TO ADDRESS THE ISSUE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1995267 | T:SLIM X2 INSULIN PUMP WITH CONTROL-IQ TECHNOLOGY | ALTERNATE CONTROLLER ENABLED INFUSION PUMP | QFG | TANDEM DIABETES CARE | 1002717 | 00850006613762 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 19 YR | Female | INFUSION SET: AUTOSOFT XC INSULIN: APIDRA |