PENTAX
Report
- Report Number
- 9610877-2022-56754
- Event Type
- Malfunction
- Date Received
- August 19, 2022
- Date of Event
- July 25, 2022
- Report Date
- August 19, 2022
- Manufacturer
- HOYA CORPORATION PENTAX TOKYO OFFICE
- Product Code
- EOB
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
THIS DEVICE IS CLASSIFIED AS IMPORT FOR EXPORT, THEREFORE 510K IS NOT APPLICABLE. MODEL VNL-1570STK IS AVAILABLE IN THE USA WITH A 510K NUMBER K162151. WE CHECKED THE RETURNED UNIT AND CONFIRMED THAT THE CCD MODULE BLACKOUT. BASED ON THE RESULT, WE CONCLUDED THAT IT WAS CAUSED DUE TO THE PHYSICAL DAMAGE APPLIED ON THE CCD MODULE. IN ADDITION, WE CONFIRMED THAT THE LIGHT GUIDE CABLE PERFORATED, THE REMOTE CONTROL BUTTONS CRACKED, THE SEGMENT CORRODED, THE CONTROL BODY CORRODED, THE CCD DRIVER PCB CORRODED, THE INSERTION FLEXIBLE TUBE (IFT) CRUSHED, AND THE LIGHT GUIDE CABLE LEAK; HOWEVER, THEY ARE NOT THE MAIN CAUSE, AND/OR IRRELEVANT TO THE ALLEGED COMPLAINT. BASED ON THE TECHNICAL REPORT ""HR-RPT-0586(IMAGE FAILURE)"" AND/OR THE RISK ANALYSIS RESULTS, IT WAS EVALUATED TO SUBMIT MDR.
THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE(BLACKOUT).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1816748 | PENTAX | NASOPHARYNGOSCOPE | EOB | HOYA CORPORATION PENTAX TOKYO OFFICE | VNL-1070STK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |