FDA Adverse Event Malfunction Summary report: N

PENTAX

MDR report key: 15256382 · Received August 19, 2022

Report

Report Number
9610877-2022-56753
Event Type
Malfunction
Date Received
August 19, 2022
Date of Event
July 25, 2022
Report Date
August 19, 2022
Manufacturer
HOYA CORPORATION PENTAX TOKYO OFFICE
Product Code
EOB
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THIS DEVICE IS CLASSIFIED AS IMPORT FOR EXPORT, THEREFORE 510K IS NOT APPLICABLE. MODEL VNL-1570STK IS AVAILABLE IN THE USA WITH A 510K NUMBER K162151. WE CHECKED THE RETURNED UNIT AND CONFIRMED THAT THE CCD MODULE BLACKOUT. BASED ON THE RESULT, WE CONCLUDED THAT IT WAS CAUSED DUE TO THE PHYSICAL DAMAGE APPLIED ON THE CCD MODULE. IN ADDITION, WE CONFIRMED THAT THE U/D KNOB BROKEN, THE LIGHT GUIDE FIBER BUNDLE (LCB) BROKEN, THE U/D LOCK LEVER BROKEN, THE REMOTE CONTROL BUTTONS CRACKED, THE CCD DRIVER PCB CORRODED, THE INSERTION FLEXIBLE TUBE (IFT) CRUSHED, THE LIGHT GUIDE CABLE FOR CONTROL BODY DEFORMED, AND THE LIGHT GUIDE CABLE FOR CONTROL BODY LEAK; HOWEVER, THEY ARE NOT THE MAIN CAUSE, AND/OR IRRELEVANT TO THE ALLEGED COMPLAINT. BASED ON THE TECHNICAL REPORT ""HR-RPT-0586(IMAGE FAILURE)"" AND/OR THE RISK ANALYSIS RESULTS, IT WAS EVALUATED TO SUBMIT MDR.

Description of Event or Problem · 0

THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE(BLACKOUT).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1816745 PENTAX NASOPHARYNGOSCOPE EOB HOYA CORPORATION PENTAX TOKYO OFFICE VNL-1070STK

Patients

Seq Age Sex Outcome Treatment
1 Unknown