BINAXNOW COVID-19 ANTIGEN SELF-TEST
Report
- Report Number
- 1221359-2022-03586
- Event Type
- Malfunction
- Date Received
- August 17, 2022
- Date of Event
- August 10, 2022
- Report Date
- August 10, 2022
- Manufacturer
- ABBOTT DIAGNOSTICS SCARBOROUGH, INC
- Product Code
- QKP
- UDI-DI
- 00811877011408
- PMA / PMN Number
- EUA210264
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
Narratives
TESTED THE BINAXNOW COVID-19 AG CARD RETAIN KIT LOT 193019 WITH COVID-19 LOD AND BLANK PATIENT SWABS. ALL TESTS WERE RUN IN TRIPLICATE, WERE VALID, AND PERFORMED AS EXPECTED. NO FALSE NEGATIVE RESULTS WERE OBSERVED, AND THE CUSTOMERS COMPLAINT WAS NOT REPLICATED. THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR KIT PART NUMBER 195-160 & 195-260 / LOT 193019 AND DEVICE PART NUMBER 195-430H / LOT 189791. QUALITY CONTROL RELEASE TESTING MET SPECIFICATIONS AND THERE WERE NO FALSE NEGATIVE RESULTS OBSERVED. THE CURRENT OVERALL INCIDENT RATE FOR FALSE NEGATIVE PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) FOR THIS SPECIFIC LOT BASED ON THE TOTAL QUANTITY OF DEVICES MANUFACTURED FOR DISTRIBUTION IS (B)(4). BASED ON THE EVIDENCE AVAILABLE, IT INDICATES THAT THIS DEVICE LOT IS PERFORMING WITHIN LABEL CLAIMS. NO FURTHER ACTION IS REQUIRED.
THE CONSUMER REPORTED A FALSE NEGATIVE RESULT WITH THE BINAXNOW COVID-19 ANTIGEN SELF TEST PERFORMED ON (B)(6) 2022 WITH A NASAL SAMPLE. PCR TESTING WAS PERFORMED ON (B)(6) 2022 AND GENERATED A POSITIVE RESULT. THE CONSUMER STATED THAT SHE WAS SYMPTOMATIC. NO ADDITIONAL PATIENT INFORMATION, INCLUDING TREATMENT AND OUTCOME, WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1950082 | BINAXNOW COVID-19 ANTIGEN SELF-TEST | LATERAL FLOW IMMUNOASSAY IVD OF COVID-19 | QKP | ABBOTT DIAGNOSTICS SCARBOROUGH, INC | 193019 | 00811877011408 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 63 YR | Female |