BD POSIFLUSH¿ PRE-FILLED SALINE SYRINGE
Report
- Report Number
- 1710034-2022-00437
- Event Type
- Malfunction
- Date Received
- August 17, 2022
- Date of Event
- July 12, 2022
- Report Date
- August 5, 2022
- Manufacturer
- BD MEDICAL (BD WEST) MEDICAL SURGICAL
- Product Code
- NGT
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- NURSE
- Health Professional
- Yes
Narratives
INVESTIGATION SUMMARY: A DEVICE HISTORY RECORD REVIEW WAS COMPLETED BY OUR QUALITY ENGINEER TEAM FOR PROVIDED MATERIAL NUMBER: 306594 AND LOT NUMBER: 1152092. THE REVIEW DID NOT REVEAL ANY DETECTED ABNORMALITIES DURING THE PRODUCTION PROCESS THAT COULD HAVE CONTRIBUTED TO THIS DEFECT AND ALL QUALITY TESTS WERE FOUND TO BE WITHIN SPECIFICATION. AS A SAMPLE WAS UNAVAILABLE FOR RETURN, A THOROUGH SAMPLE INVESTIGATION COULD NOT BE COMPLETED. BASED ON THE INVESTIGATION RESULTS, AN EXACT CAUSE FOR THIS INCIDENT COULD NOT BE IDENTIFIED. THERE ARE QUALITY CONTROLS CURRENTLY IN PLACE TO DETECT THIS TYPE OF DEFECT DURING THE PRODUCTION PROCESS. FURTHER ACTION HAS NOT BEEN DETERMINED NECESSARY AT THIS TIME. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. OUR QUALITY TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.
IT WAS REPORTED THAT THE PLUNGER ROD BROKE ON THE BD POSIFLUSH¿ PRE-FILLED SALINE SYRINGE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: (B)(6) 2022 WHEN THE NURSE OPENED THE NEWLY PACKAGED FLUSH IN THE TREATMENT ROOM OF THE FOURTH AREA, SHE FOUND THAT THE TUBE HANDLE WAS BROKEN AND COULD NOT USE THE FLUSH NORMALLY, AND IMMEDIATELY REPLACED IT WITH A NEW ONE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2013659 | BD POSIFLUSH¿ PRE-FILLED SALINE SYRINGE | SALINE VASCULAR ACCESS FLUSH | NGT | BD MEDICAL (BD WEST) MEDICAL SURGICAL | 1152092 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |