FDA Adverse Event Malfunction Summary report: N

CD HORIZON SPINAL SYSTEM

MDR report key: 15238397 · Received August 17, 2022

Report

Report Number
1030489-2022-00778
Event Type
Malfunction
Date Received
August 17, 2022
Date of Event
July 1, 2022
Report Date
December 12, 2022
Manufacturer
WARSAW ORTHOPEDICS
Product Code
KWP
PMA / PMN Number
SEE H10
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THIS PART IS NOT APPROVED FOR SALE IN UNITED STATES; HOWEVER, A LIKE DEVICE WITH CATALOG# 1556006300, 510(K)# K121680 AND UDI# (B)(4), IS APPROVED FOR SALE IN UNITED STATES. MEDTRONIC IS SUBMITTING THIS REPORT TO COMPLY WITH FDA REPORTING REGULATIONS UNDER 21 CFR PARTS 4 AND 803. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION AND HAS PROVIDED AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY ¿DEFECTS¿ OR HAS ¿MALFUNCTIONED¿. THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REGULATORY REPORTING. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE THEM BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.

Additional Manufacturer Narrative · 0

H6:PRODUCT ANALYSIS PART#:1555106300 LOT# UNKNOWN THERE WERE TWO RODS RETURNED BOTH RODS HAVE BROKEN IN A SIMILAR MODE. NO PRE-EXISTING SURFACE DEFECTS WERE IDENTIFIED THAT COULD CONTRIBUTE TO THE BREAKS. A MICROSCOPIC REVIEW OF THE FRACTURE SURFACE REVEALS AN AREA OF CRACK PROPAGATION WITH A SHEER LIP ON THE OTHER SIDE OF THE FRACTURE FACE. THERE DID NOT APPEAR TO BE ANY EVIDENCE OF CYCLIC FATIGUE ACROSS THE FRACTURE FACE. BOTH THE AREA OF CRACK PROPAGATION AND THE SHEER LIP ARE IN PLANE WITH THE LOCATION OF THE SET SCREW WITNESS MARK. THE PREVIOUS OBSERVATIONS ARE CONSISTENT WITH BEND STRESS OVERLOAD. MEDTRONIC IS SUBMITTING THIS REPORT TO COMPLY WITH FDA REPORTING REGULATIONS UNDER 21 CFR PARTS 4 AND 803. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION AND HAS PROVIDED AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY ¿DEFECTS¿ OR HAS ¿MALFUNCTIONED¿. THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REGULATORY REPORTING. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE THEM BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 0

INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING SPINAL THERAPY FOR BONE FUSION FAILURE. IT WAS REPORTED THAT THE ROD WAS BROKEN. ADDITIONAL SURGERY PREFORMED. REPLACEMENT AND REINFORCEMENT OF ROD WAS DONE. PRODUCT WILL BE RETURNED. ADDITIONAL INFORMATION RECEIVED FROM MANUFACTURER REPRESENTATIVE THAT THE INITIAL SURGERY WAS ON (B)(6) 2019. THE PROCEDURE INVOLVED WAS PSF FOR PURULENT SPONDYLITIS. IT IS UNKNOWN WHETHER THERE WAS ANY PATIENT COMPLICATION AS A RESULT OF THIS EVENT. ADDITIONAL INFORMATION RECEIVED FROM MANUFACTURER REPRESENTATIVE THAT TWO RODS WERE BROKEN. THE OTHER ROD WAS REMOVED ALONG WITH THE REPORTED RODS, BUT IT HAS NEITHER MALFUNCTIONING NOR ALLEGATION. THE CONNECTORS ALSO NOT MALFUNCTIONED, THEY ARE SIMPLY CONNECTED TO THE BROKEN RODS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1916140 CD HORIZON SPINAL SYSTEM APPLIANCE, FIXATION, SPINAL INTERLAMINAL KWP WARSAW ORTHOPEDICS 1555106300 UNKNOWN

Patients

Seq Age Sex Outcome Treatment
1 75 YR Male Required Intervention