FDA Adverse Event Malfunction Summary report: N

BD INTIMA-II¿ CLOSED IV CATHETER SYSTEM

MDR report key: 15236944 · Received August 16, 2022

Report

Report Number
3006948883-2022-00168
Event Type
Malfunction
Date Received
August 16, 2022
Date of Event
July 5, 2022
Report Date
August 9, 2022
Manufacturer
BD (SUZHOU)
Product Code
FOZ
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION SUMMARY: IN RESPONSE TO THE EVENT REPORTED BY THE FACILITY A DEVICE HISTORY REVIEW WAS CONDUCTED FOR LOT NUMBER 2011052. OUR RECORDS SHOW THAT THIS IS THE ONLY INSTANCE OF THIS ISSUE OCCURRING IN THIS PRODUCTION BATCH. ACCORDING TO THE SAMPLING PLAN APPLIED FOR PRODUCT PERFORMANCE, THIS LOT WAS ACCEPTED AND RELEASED WITHOUT DEFECTS BEING NOTED DURING THE FINAL ASSEMBLY OR VISUAL INSPECTIONS. A SAMPLE COULD NOT BE OBTAINED FOR EVALUATION AND TESTING; IN LIEU OF THE AFFECTED DEVICE, FUNCTIONAL TESTING WAS PERFORMED ON RETENTION SAMPLES FOR THIS LOT, THE RESULTS OF THESE SHOW THAT THE TESTED UNITS PERFORMED WITHIN PRODUCT SPECIFICATIONS. UNFORTUNATELY WITHOUT THE ABILITY TO INVESTIGATE THE AFFECTED UNIT OUR QUALITY ENGINEERS WERE UNABLE TO DETERMINE THE ROOT CAUSE FOR THIS COMPLAINT.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE BD INTIMA-II¿ CLOSED IV CATHETER SYSTEM LEAKED FROM THE NEEDLE CORE. AFTER THE SUCCESSFUL PUNCTURE THE NEEDLE CORE WAS PULLED OUT, THE DRUG FLUID SPILT. AFTER IDENTIFY PATIENT WITHOUT HISTORY OF INFECTION AT THE SAME TIME, OTHERS IMMEDIATELY WASHED SPILT AREA FOR NURSE WITH A LOT OF SALT WATER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1223194 BD INTIMA-II¿ CLOSED IV CATHETER SYSTEM INTRAVASCULAR CATHETER FOZ BD (SUZHOU) 2011052

Patients

Seq Age Sex Outcome Treatment
1 Unknown