BD PEGASUS¿ SAFETY CLOSED IV CATHETER SYSTEM
Report
- Report Number
- 3014704491-2022-00355
- Event Type
- Malfunction
- Date Received
- August 16, 2022
- Date of Event
- June 24, 2022
- Report Date
- September 1, 2022
- Manufacturer
- BD (SUZHOU)
- Product Code
- FOZ
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- NURSE
- Health Professional
- Yes
Narratives
A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
CORRECTION: H1: DEVICE EVAL BY MANUFACTURER?: NO. H6: INVESTIGATION SUMMARY: A DEVICE HISTORY REVIEW WAS CONDUCTED FOR LOT NUMBER 1053329. OUR RECORDS SHOW THAT THIS IS THE ONLY INSTANCE OF THIS ISSUE OCCURRING IN THIS PRODUCTION BATCH. ACCORDING TO THE SAMPLING PLAN APPLIED FOR PRODUCT PERFORMANCE, THIS LOT WAS ACCEPTED AND RELEASED WITHOUT DEFECTS BEING NOTED DURING THE FINAL ASSEMBLY OR VISUAL INSPECTIONS. A SAMPLE COULD NOT BE OBTAINED FOR EVALUATION AND TESTING; IN LIEU OF THE AFFECTED DEVICE, VISUAL ANALYSIS WAS PERFORMED ON RETENTION SAMPLES FOR THIS LOT, THE RETAINED DEVICES WERE FOUND TO BE FREE OF ANY DAMAGE OR OTHER ABNORMALITIES. UNFORTUNATELY WITHOUT THE ABILITY TO INVESTIGATE THE AFFECTED UNIT OUR QUALITY ENGINEERS WERE UNABLE TO DETERMINE THE ROOT CAUSE FOR THIS COMPLAINT.
IT WAS REPORTED THAT THERE WAS WHITE FOREIGN MATTER FOUND IN THE CATHETER TUBE OF THE INDWELLING OF THE BD PEGASUS¿ SAFETY CLOSED IV CATHETER SYSTEM BEFORE STARTING THE INFUSION. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: THE NURSE FOUND WHITE FLOC IN THE CATHETER TUBE OF INDWELLING NEEDLE WHEN PREPARING MATERIALS FOR THE TREATMENT OF THE PATIENT, AND IMMEDIATELY REPLACED IT.
IT WAS REPORTED THAT THERE WAS WHITE FOREIGN MATTER FOUND IN THE CATHETER TUBE OF THE INDWELLING OF THE BD PEGASUS¿ SAFETY CLOSED IV CATHETER SYSTEM BEFORE STARTING THE INFUSION. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: THE NURSE FOUND WHITE FLOC IN THE CATHETER TUBE OF INDWELLING NEEDLE WHEN PREPARING MATERIALS FOR THE TREATMENT OF THE PATIENT, AND IMMEDIATELY REPLACED IT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1509657 | BD PEGASUS¿ SAFETY CLOSED IV CATHETER SYSTEM | INTRAVASCULAR CATHETER | FOZ | BD (SUZHOU) | 1053329 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |