ZOLL IVTM THERMOGARD XP
Report
- Report Number
- 3010617000-2022-01346
- Event Type
- Malfunction
- Date Received
- August 16, 2022
- Date of Event
- July 13, 2022
- Report Date
- August 16, 2022
- Manufacturer
- ZOLL CIRCULATION INC
- Product Code
- NCX
- UDI-DI
- 00849111075022
- PMA / PMN Number
- K072234
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
ZOLL HAS NOT YET RECEIVED THE DEVICE IN THE COMPLAINT FOR INVESTIGATION. A SUPPLEMENTAL REPORT WILL BE FILED IF AND WHEN THE PRODUCT IS RETURNED AND INVESTIGATION HAS BEEN COMPLETED.
IVTM THERAPY WAS STARTED FOR A NEURO PATIENT TO COOL THEM. AFTER A FEW HOURS, THE CUSTOMER NOTICED THE VOLUME OF THE SALINE BAG (500ML) WAS DECREASING. DURING THE NIGHT, THE THERMOGARD XP IVTM SYSTEM (SN UNKNOWN) DISPLAYED AN AIR TRAP ALARM AND THERE WAS SALINE OBSERVED IN THE AIR TRAP HOLDER AND AROUND THE CONSOLE. NO BLOOD WAS OBSERVED IN THE TUBING OF THE START-UP KIT (SUK) (LOT #UNKNOWN) AND THE LOCATION OF THE LEAK IS UNKNOWN. IT IS UNKNOWN IF THE ALARM WAS CLEARED. THE CUSTOMER REPLACED THE SALINE BAG AND SUK AND SWITCHED THE THERMOGARD XP IVTM SYSTEM, BUT THE SALINE BAG STILL DEPLETED. THE CUSTOMER SUSPECTS THE ICY CATHETER (LOT #170133) HAS A LEAK. NO DEVICE MALFUNCTION WAS REPORTED WITH THE SECOND SUK OR CONSOLE. THE PATIENT WAS DISCONNECTED FROM THE THERMOGARD AND ARCTIC SUN WAS USED INSTEAD. UPON FOLLOW UP, THE CUSTOMER REPORTED THE SUK AND CATHETER WERE DISCARDED AND WILL NOT BE RETURNED. NO CONSEQUENCE OR IMPACT TO THE PATIENT. PLEASE SEE THE FOLLOWING RELATED MFR REPORT: PLEASE SEE THE FOLLOWING RELATED MFR REPORT:MFR #3010617000-2022-01112 FOR THE ICY CATHETER
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2013137 | ZOLL IVTM THERMOGARD XP | THERMAL REGULATING SYSTEM | NCX | ZOLL CIRCULATION INC | 8700-0650 | 00849111075022 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |